Rack-and-Pinion Displacement of an Intramedullary Pin During Minimally Invasive Plate–Rod Osteosynthesis of the Canine Femur—A Case Report

We present a case of unexpected proximal displacement of an intramedullary pin (IMP) during plate-rod repair of a femoral fracture caused by minimally invasive plate osteosynthesis (MIPO), requiring immediate revision. Implant retrieval analysis and ex vivo modelling were performed to characterise the technique failure mode. The case details are reported. Implant retrieval analysis consisted of stereo zoom microscopic examination of the retrieved IMP. Wear patterns formed by conflict with a 2.8 mm, two-fluted surgical drill bit and a 3.5 mm AO locking screw were replicated using a simple paper impression model. The mechanism of pin movement was replicated in a benchtop laminated polyurethane foam block model, and wear patterns produced during drilling and screw insertion were characterised using stereo zoom. The wear pattern visible on the retrieved IMP suggested axial displacement caused by a rack-and-pinion-like mechanism, enacted by contact with either the drill bit or locking screws during placement of the repair construct. Significant axial displacement of the IMP due to conflict with screws during construct placement is possible during the placement of plate-rod fixation. Surgeons should confirm implant positioning if implant conflict is recognised intra-operatively.

Authors

Institutions

Publication Details

Journal
Animals
Published
2025-09-23
DOI
https://doi.org/10.3390/ani15192777
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Rack-and-Pinion Displacement of an Intramedullary Pin During Minimally Invasive Plate–Rod Osteosynthesis of the Canine Femur—A Case Report

William Robert Walsh, Dan Wills, Max J. Lloyd, Kristy L. Hospes
Animals
Orthopaedic implants and arthroplasty
article

Rack-and-Pinion Displacement of an Intramedullary Pin During Minimally Invasive Plate–Rod Osteosynthesis of the Canine Femur—A Case Report

William Robert Walsh, Dan Wills, Max J. Lloyd, Kristy L. Hospes
article en

Abstract

We present a case of unexpected proximal displacement of an intramedullary pin (IMP) during plate-rod repair of a femoral fracture caused by minimally invasive plate osteosynthesis (MIPO), requiring immediate revision. Implant retrieval analysis and ex vivo modelling were performed to characterise the technique failure mode. The case details are reported. Implant retrieval analysis consisted of stereo zoom microscopic examination of the retrieved IMP. Wear patterns formed by conflict with a 2.8 mm, two-fluted surgical drill bit and a 3.5 mm AO locking screw were replicated using a simple paper impression model. The mechanism of pin movement was replicated in a benchtop laminated polyurethane foam block model, and wear patterns produced during drilling and screw insertion were characterised using stereo zoom. The wear pattern visible on the retrieved IMP suggested axial displacement caused by a rack-and-pinion-like mechanism, enacted by contact with either the drill bit or locking screws during placement of the repair construct. Significant axial displacement of the IMP due to conflict with screws during construct placement is possible during the placement of plate-rod fixation. Surgeons should confirm implant positioning if implant conflict is recognised intra-operatively.

AnimalsVol. 15(19)
Central Coast Local Health District (AU), Sydney Orthopaedic Research Institute (AU)
Openalex Percentile: Top 59%
Orthopaedic implants and arthroplasty
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Rack-and-Pinion Displacement of an Intramedullary Pin During Minimally Invasive Plate–Rod Osteosynthesis of the Canine Femur—A Case Report — William Robert Walsh, Dan Wills, et al. · Animals (2025) | TGRS Research Map | TGRS