Comparison of multifocal electroretinography results with two different electrodes in patients using hydroxychloroquine: a preliminary study

Purpose:To evaluate retinal electrophysiological responses with two different electrodes in patients who have been using HCQ at a dose of least 200 mg/day for more than five years and who are not known to have hydroxychloroquine (HCQ) retinal toxicity and to compare JET electrode data and H-K loop electrode data.Method: 20 eyes of 20 participants, 10 patients and 10 controls to whom HCQ treatment will be started, who were admitted consecutively to the clinic between October 2020 and March 2021 and who met the inclusion criteria were evaluated prospectively.Results: When the P1 wave amplitudes of both electrodes in the patient and control groups were compared, it was obtained lower in all waves in the patient group; Statistically significant difference was observed only in Ring 1 (p=0.029) in JET electrode, in ring 2 (p=0.019) and ring 3 (p=0.009) in H-K loop electrode. Discussion:The reason for the variation of the affected ring with respect to the electrode may be associated with the recorded localization of the electrode.In this respect, internal consistency of the electrodes draws attention.

Authors

Publication Details

Journal
Journal of Retina-Vitreous
Published
2025-01-01
DOI
https://doi.org/10.37845/ret.vit.2025.34.28
Primary Topic
Drug-Induced Ocular Toxicity
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of multifocal electroretinography results with two different electrodes in patients using hydroxychloroquine: a preliminary study

İnci Güngör, Mustafa Subaşı
Journal of Retina-Vitreous
Drug-Induced Ocular Toxicity
article

Comparison of multifocal electroretinography results with two different electrodes in patients using hydroxychloroquine: a preliminary study

İnci Güngör, Mustafa Subaşı
article en

Abstract

Purpose:To evaluate retinal electrophysiological responses with two different electrodes in patients who have been using HCQ at a dose of least 200 mg/day for more than five years and who are not known to have hydroxychloroquine (HCQ) retinal toxicity and to compare JET electrode data and H-K loop electrode data.Method: 20 eyes of 20 participants, 10 patients and 10 controls to whom HCQ treatment will be started, who were admitted consecutively to the clinic between October 2020 and March 2021 and who met the inclusion criteria were evaluated prospectively.Results: When the P1 wave amplitudes of both electrodes in the patient and control groups were compared, it was obtained lower in all waves in the patient group; Statistically significant difference was observed only in Ring 1 (p=0.029) in JET electrode, in ring 2 (p=0.019) and ring 3 (p=0.009) in H-K loop electrode. Discussion:The reason for the variation of the affected ring with respect to the electrode may be associated with the recorded localization of the electrode.In this respect, internal consistency of the electrodes draws attention.

Journal of Retina-VitreousVol. 34(3)
Openalex Percentile: Top 62%
Drug-Induced Ocular Toxicity
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Comparison of multifocal electroretinography results with two different electrodes in patients using hydroxychloroquine: a preliminary study — İnci Güngör, Mustafa Subaşı · Journal of Retina-Vitreous (2025) | TGRS Research Map | TGRS