The TR5500 Thermoluminescent Dosimeter (TLD) reader was specifically designed and developed for the automated and efficient readout of Thermoluminescent (TL) chips and pellets in Iran. In accordance with International Atomic Energy Agency (IAEA) requirements for dosimetry laboratories, implementing a comprehensive Quality Management System (QMS) is essential to ensure adherence to standard criteria. This study aimed to evaluate key characteristics of the reader, including response non-linearity, Coefficients of Variation (COV), readout stability over time, and the reproducibility of TLD measurements. LiF: Mg, Cu, P TLDs were exposed to a standard photon radiation field and subsequently read using the TR5500 reader. The characteristics were then calculated and compared against the relevant criteria outlined in International Electrotechnical Commission (IEC) and American Society for Testing and Materials (ASTM) standards, such as ASTM E668-10 and IEC 62387. The reader met the key international standard requirements and, in certain aspects, even surpassed them. The device exhibited excellent stability and reproducibility in readouts, confirming its robust reliability and long-term suitability for routine use in various accredited dosimetry laboratories.

Authors

Institutions

Publication Details

Journal
Journal of Biomedical Physics and Engineering
Published
2025-01-01
DOI
https://doi.org/10.31661/jbpe.v0i0.2506-1937
Primary Topic
VLSI and Analog Circuit Testing
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Performance Characteristics of a New Automatic Chip TLD Reader

S.M. Hosseini Pooya, Sajjad Betyar, Meysam Divsalar
Journal of Biomedical Physics and Engineering
VLSI and Analog Circuit Testing
article

Performance Characteristics of a New Automatic Chip TLD Reader

S.M. Hosseini Pooya, Sajjad Betyar, Meysam Divsalar
article en

Abstract

The TR5500 Thermoluminescent Dosimeter (TLD) reader was specifically designed and developed for the automated and efficient readout of Thermoluminescent (TL) chips and pellets in Iran. In accordance with International Atomic Energy Agency (IAEA) requirements for dosimetry laboratories, implementing a comprehensive Quality Management System (QMS) is essential to ensure adherence to standard criteria. This study aimed to evaluate key characteristics of the reader, including response non-linearity, Coefficients of Variation (COV), readout stability over time, and the reproducibility of TLD measurements. LiF: Mg, Cu, P TLDs were exposed to a standard photon radiation field and subsequently read using the TR5500 reader. The characteristics were then calculated and compared against the relevant criteria outlined in International Electrotechnical Commission (IEC) and American Society for Testing and Materials (ASTM) standards, such as ASTM E668-10 and IEC 62387. The reader met the key international standard requirements and, in certain aspects, even surpassed them. The device exhibited excellent stability and reproducibility in readouts, confirming its robust reliability and long-term suitability for routine use in various accredited dosimetry laboratories.

Journal of Biomedical Physics and EngineeringVol. online(5)
Shahed University (IR)
Openalex Percentile: Top 72%
VLSI and Analog Circuit Testing
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.