Computational system for extremity dosimetry in nuclear medicine
The goal of this research is to create a computational dosimetry system to monitor the dose received by nuclear medicine workers on their extremities. The system creates multiple Monte Carlo simulations to assess the dose received in different sections of the hands. Each simulation is created based on depth camera images of the manipulation of radioactive sources, using digital twins to represent the hands of the worker and the radiation source. The position of the hands and the source is constantly recalculated using computer vision algorithms. In this paper, we explain the different components of the computational system, assessing the possible sources of uncertainty introduced by each component. As a proof of concept, we compared the dose estimated by the computational system with real dosimeter measurements for two different scenarios. The simulated dose was, on average, 51% of the measured dose.
Authors
- Mahmoud Abdelrahman (ORCID: https://orcid.org/0000-0002-0774-3478)
- Lara Struelens
- F. Vanhavere (ORCID: https://orcid.org/0000-0003-3813-0858)
- Pasquale Lombardo (ORCID: https://orcid.org/0000-0001-7095-7295)
- Werner Schoonjans
- Niki Bergans (ORCID: https://orcid.org/0000-0003-4927-5315)
- Daniel Santiago Rondón (ORCID: https://orcid.org/0009-0000-3342-0093)
Institutions
- Belgian Nuclear Research Centre (BE)
- KU Leuven (BE)
Publication Details
- Journal
- Journal of Radiological Protection
- Published
- 2025-09-23
- DOI
- https://doi.org/10.1088/1361-6498/ae0a5b
- Primary Topic
- Radiation Dose and Imaging
- Type
- article
- Field-Weighted Citation Impact
- 0.00