Intracuff Water Droplet Accumulation Occurring Around Eight Hours in Polyurethane-Cuffed Endotracheal Tubes: An In Vitro Simulation Study

Introduction Ventilator-associated pneumonia (VAP) is a major complication in intensive care unit (ICU) patients, often resulting from microaspiration due to poor cuff sealing. Polyurethane (PU)-cuffed tracheal tubes, with thinner walls than conventional polyvinyl chloride (PVC) cuffs, offer improved sealing. However, their high water permeability can lead to intracuff water droplet accumulation, potentially interfering with accurate cuff pressure monitoring. This study aimed to determine the timing of water droplet formation in PU cuffs under simulated ventilatory conditions. Methods An in vitro simulation was conducted using six tracheal tubes (two each of PU-cuffed Microcuff™ (Kimberly-Clark Health Care, Roswell, GA), PU-cuffed Parker Flex-Tip™ (Parker Medical, Highlands Ranch, CO), and PVC-cuffed TaperGuard™ (Covidien (Medtronic), Mansfield, MA)). Tubes were exposed to heated, humidified airflow for 24 hours at a mean temperature of 39.8 ± 0.4°C and 88 ± 4.6% humidity. Cuff pressure was maintained at 25 cmH₂O. Hourly inspections for visible intracuff water droplets were performed. Results Visible water droplets first appeared at eight hours in all PU-cuffed tubes, with a mean onset of 8.25 hours. Fluid migration into the inflation line and pilot balloon was also observed. No water accumulation was seen in the PVC-cuffed tubes. No structural cuff damage was noted. Conclusion Water droplet accumulation in PU-cuffed tracheal tubes occurs around eight hours after exposure to humidified ventilation. This early condensation may impair cuff pressure accuracy, posing a safety risk in ICU settings. Clinicians should consider this when managing airway devices and cuff pressures in patients with PU-cuffed tubes.

Authors

Institutions

Publication Details

Journal
Cureus
Published
2025-09-23
DOI
https://doi.org/10.7759/cureus.93000
Primary Topic
Nosocomial Infections in ICU
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Intracuff Water Droplet Accumulation Occurring Around Eight Hours in Polyurethane-Cuffed Endotracheal Tubes: An In Vitro Simulation Study

Nobuyuki Yokoyama, Masafumi Idei, Yohei Sakai, Makoto Inoue et al.
Cureus
Nosocomial Infections in ICU
article

Intracuff Water Droplet Accumulation Occurring Around Eight Hours in Polyurethane-Cuffed Endotracheal Tubes: An In Vitro Simulation Study

Nobuyuki Yokoyama, Masafumi Idei, Yohei Sakai, Makoto Inoue, Naoto Funami
article en

Abstract

Introduction Ventilator-associated pneumonia (VAP) is a major complication in intensive care unit (ICU) patients, often resulting from microaspiration due to poor cuff sealing. Polyurethane (PU)-cuffed tracheal tubes, with thinner walls than conventional polyvinyl chloride (PVC) cuffs, offer improved sealing. However, their high water permeability can lead to intracuff water droplet accumulation, potentially interfering with accurate cuff pressure monitoring. This study aimed to determine the timing of water droplet formation in PU cuffs under simulated ventilatory conditions. Methods An in vitro simulation was conducted using six tracheal tubes (two each of PU-cuffed Microcuff™ (Kimberly-Clark Health Care, Roswell, GA), PU-cuffed Parker Flex-Tip™ (Parker Medical, Highlands Ranch, CO), and PVC-cuffed TaperGuard™ (Covidien (Medtronic), Mansfield, MA)). Tubes were exposed to heated, humidified airflow for 24 hours at a mean temperature of 39.8 ± 0.4°C and 88 ± 4.6% humidity. Cuff pressure was maintained at 25 cmH₂O. Hourly inspections for visible intracuff water droplets were performed. Results Visible water droplets first appeared at eight hours in all PU-cuffed tubes, with a mean onset of 8.25 hours. Fluid migration into the inflation line and pilot balloon was also observed. No water accumulation was seen in the PVC-cuffed tubes. No structural cuff damage was noted. Conclusion Water droplet accumulation in PU-cuffed tracheal tubes occurs around eight hours after exposure to humidified ventilation. This early condensation may impair cuff pressure accuracy, posing a safety risk in ICU settings. Clinicians should consider this when managing airway devices and cuff pressures in patients with PU-cuffed tubes.

CureusVol. 17(9)
Yokohama City University Hospital (JP), Yokohama City University (JP)
Openalex Percentile: Top 70%
Nosocomial Infections in ICU
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.