Radiation dose reduction in abdominopelvic CT using BMI specific low kV protocol

To evaluate the impact of a body mass index (BMI)-adjusted low kV protocol on radiation dose reduction and image quality in contrast-enhanced computed tomography (CECT) scans of abdomen and pelvis. Among the patients referred for CECT abdomen and pelvis scans, 120 were scanned using the standard protocol at 120 kV, whereas 96 were scanned using a BMI-specific low-radiation dose protocol. Patients were categorized into 3 BMI groups, low BMI (<18.5 kg/m 2 ) scanned at 80 kV, and normal BMI (18.5–24.9 kg/m 2 ) and high BMI (25–29.9 kg/m 2 ) at 100 kV. Effective dose ( E ) was recorded for each group. Image quality was assessed in the arterial and portal venous series. Quantitative image quality analysis was performed to estimate the signal-to-noise ratio (SNR), contrast to noise ratio (CNR), contrast enhancement index (CEI) and image noise. Qualitative analysis was independently assessed using a 5-point scale. The study reported higher radiation doses with higher BMI, however, low dose protocol achieved significant reduction in radiation dose across all BMI groups: low BMI, (21± 2) vs ( 10 ± 1.7) mSv ( P < 0.05); normal BMI, (22 ± 2.3) vs (14 ± 1.3) mSv ( P < 0.05); high BMI, (25 ± 2.8) vs (17 ± 1.8) mSv ( P < 0.05), for the standard and low dose protocols, respectively. Low dose protocols reported to have a higher image noise and CEI, lower SNR and comparable CNR across BMI groups in both arterial and portal venous series. Qualitative scores remain within acceptable diagnostic limits across BMI groups. BMI-specific low-kV CT protocols significantly reduced the radiation dose across all BMI categories while maintaining acceptable diagnostic image quality, suggesting that tailored low dose scanning may be feasible for clinical use.

Authors

Institutions

Publication Details

Journal
Radiation Medicine and Protection
Published
2025-09-23
DOI
https://doi.org/10.1016/j.radmp.2025.09.006
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Radiation dose reduction in abdominopelvic CT using BMI specific low kV protocol

Nitika C. Panakkal, Suresh Sukumar, Rajagopal Kadavigere, N Ravishankar
Radiation Medicine and Protection
Radiation Dose and Imaging
article

Radiation dose reduction in abdominopelvic CT using BMI specific low kV protocol

Nitika C. Panakkal, Suresh Sukumar, Rajagopal Kadavigere, N Ravishankar
article en

Abstract

To evaluate the impact of a body mass index (BMI)-adjusted low kV protocol on radiation dose reduction and image quality in contrast-enhanced computed tomography (CECT) scans of abdomen and pelvis. Among the patients referred for CECT abdomen and pelvis scans, 120 were scanned using the standard protocol at 120 kV, whereas 96 were scanned using a BMI-specific low-radiation dose protocol. Patients were categorized into 3 BMI groups, low BMI (<18.5 kg/m 2 ) scanned at 80 kV, and normal BMI (18.5–24.9 kg/m 2 ) and high BMI (25–29.9 kg/m 2 ) at 100 kV. Effective dose ( E ) was recorded for each group. Image quality was assessed in the arterial and portal venous series. Quantitative image quality analysis was performed to estimate the signal-to-noise ratio (SNR), contrast to noise ratio (CNR), contrast enhancement index (CEI) and image noise. Qualitative analysis was independently assessed using a 5-point scale. The study reported higher radiation doses with higher BMI, however, low dose protocol achieved significant reduction in radiation dose across all BMI groups: low BMI, (21± 2) vs ( 10 ± 1.7) mSv ( P < 0.05); normal BMI, (22 ± 2.3) vs (14 ± 1.3) mSv ( P < 0.05); high BMI, (25 ± 2.8) vs (17 ± 1.8) mSv ( P < 0.05), for the standard and low dose protocols, respectively. Low dose protocols reported to have a higher image noise and CEI, lower SNR and comparable CNR across BMI groups in both arterial and portal venous series. Qualitative scores remain within acceptable diagnostic limits across BMI groups. BMI-specific low-kV CT protocols significantly reduced the radiation dose across all BMI categories while maintaining acceptable diagnostic image quality, suggesting that tailored low dose scanning may be feasible for clinical use.

Radiation Medicine and ProtectionVol. 6(5)
University of Delhi (IN), Manipal Academy of Higher Education (IN)
Openalex Percentile: Top 62%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.