Comparison of Ocular Wavefront in Seated and Supine Positions Using a Hand-Held Hartmann–Shack Aberrometer

<b>Background/Objectives</b>: Intraoperative aberrometry has gained clinical relevance in correcting aberrations during cataract and corneal refractive surgeries. As wavefront aberrations are typically measured with patients seated, while surgeries are performed supine, this study aimed to compare ocular aberrations between seated and supine positions, using a hand-held Hartmann-Shack aberrometer. <b>Methods</b>: Total ocular wavefront aberrations were measured five times consecutively from cyclopledged eyes for a 6 mm pupil, under three conditions: (1) seated with a holder-fixed mode (reference), (2) seated with a hand-held mode, and (3) supine with a hand-held mode. Condition 2 was included to assess potential measurement errors from the hand-held mode. Repeatability was assessed using the standard deviation of repeated measurements (Srm) and the intraclass correlation coefficient (ICC). Differences among the three conditions were analyzed to evaluate the effects of positional change and hand-held stability on ocular wavefront aberration measurements. <b>Results</b>: Eighteen healthy subjects (36 eyes) were enrolled. The Srms for the sphere were 0.37, 0.38, and 0.40 diopters (D); and for the cylinder, 0.10, 0.10, and 0.11 D, with no significant differences across conditions. ICC values exceeded 0.9 for both lower-order aberrations (LOAs) and higher-order aberrations (HOAs), indicating excellent repeatability. The mean root mean square HOAs (HOA_RMS) were 0.48, 0.49, and 0.45 µm, with no statistically significant differences by position (<i>p</i> = 0.913) or measurement mode (<i>p</i> = 0.966). <b>Conclusions</b>: The hand-held Hartmann-Shack aberrometer demonstrated satisfactory repeatability for LOAs and HOAs. Supine measurements did not differ from seated, supporting the feasibility of extending preoperative results to intraoperative aberrometry, regardless of positional differences.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Medicine
Published
2025-09-22
DOI
https://doi.org/10.3390/jcm14186688
Citations
2
Primary Topic
Intraocular Surgery and Lenses
Type
article
Field-Weighted Citation Impact
3.41

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of Ocular Wavefront in Seated and Supine Positions Using a Hand-Held Hartmann–Shack Aberrometer

Jong Hwa Jun, Seung Pil Bang, Nicolas S. Brown, Noh Eun Kwon
2 citations
Journal of Clinical Medicine
Intraocular Surgery and Lenses
3.41
article

Comparison of Ocular Wavefront in Seated and Supine Positions Using a Hand-Held Hartmann–Shack Aberrometer

Jong Hwa Jun, Seung Pil Bang, Nicolas S. Brown, Noh Eun Kwon
article en
2 citations

Abstract

<b>Background/Objectives</b>: Intraoperative aberrometry has gained clinical relevance in correcting aberrations during cataract and corneal refractive surgeries. As wavefront aberrations are typically measured with patients seated, while surgeries are performed supine, this study aimed to compare ocular aberrations between seated and supine positions, using a hand-held Hartmann-Shack aberrometer. <b>Methods</b>: Total ocular wavefront aberrations were measured five times consecutively from cyclopledged eyes for a 6 mm pupil, under three conditions: (1) seated with a holder-fixed mode (reference), (2) seated with a hand-held mode, and (3) supine with a hand-held mode. Condition 2 was included to assess potential measurement errors from the hand-held mode. Repeatability was assessed using the standard deviation of repeated measurements (Srm) and the intraclass correlation coefficient (ICC). Differences among the three conditions were analyzed to evaluate the effects of positional change and hand-held stability on ocular wavefront aberration measurements. <b>Results</b>: Eighteen healthy subjects (36 eyes) were enrolled. The Srms for the sphere were 0.37, 0.38, and 0.40 diopters (D); and for the cylinder, 0.10, 0.10, and 0.11 D, with no significant differences across conditions. ICC values exceeded 0.9 for both lower-order aberrations (LOAs) and higher-order aberrations (HOAs), indicating excellent repeatability. The mean root mean square HOAs (HOA_RMS) were 0.48, 0.49, and 0.45 µm, with no statistically significant differences by position (<i>p</i> = 0.913) or measurement mode (<i>p</i> = 0.966). <b>Conclusions</b>: The hand-held Hartmann-Shack aberrometer demonstrated satisfactory repeatability for LOAs and HOAs. Supine measurements did not differ from seated, supporting the feasibility of extending preoperative results to intraoperative aberrometry, regardless of positional differences.

Journal of Clinical MedicineVol. 14(18)
Keimyung University Dongsan Medical Center (KR), Keimyung University (KR), University of Rochester (US)
National Research Foundation, National Research Foundation of Korea
Openalex Percentile: Top 7%
Intraocular Surgery and Lenses
3.41
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.