LC-MS/MS method for quantification of avacopan in human plasma from patients treated for antineutrophil cytoplasmic antibody-associated vasculitis

Avacopan is a new treatment for antineutrophil cytoplasmic antibody-associated vasculitis (ANCA-AAV). To date, there has been a lack of published bioanalytical methods to assay it in plasma, resulting in sparse pharmacokinetic data in clinical settings. The objective of this study was to develop and validate a liquid chromatography tandem mass spectrometry (LC-MS/MS) method for the quantification of avacopan in human plasma. [<sup>2</sup>H<sub>4</sub>]-Avacopan was as used as internal standard (IS). Samples were prepared by protein precipitation and separated was on an Accurore® C18 column (2.1 ×50 mm, 2.6 µm), with an elution gradient at a flow rate of 0.5 mL/min. The mobile phase consisted of acetonitrile (0.1 % formic acid) and water (0.1 % formic acid). The analysis run time was 6 min. Avacopan was detected by electrospray ionization on a TSQ Quantis® triple quadrupole mass spectrometer (ThermoFisher Scientific). The linearity of method ranged from 10 to 800 ng/mL. The within-run and between-run relative standard deviations were < 10.2 %. The within-run and between-run relative errors ranged from 2.4 % to 14.4 %. The IS-normalized matrix effect ranged from 2.2 % to 5.1 %, and the IS-normalized extraction recovery ranged from 104.3 % to 109.7 %. The method was fully validated, including checks on linearity, dilution integrity and carry-over. Plasma samples from 16 patients undergoing treatment for ANCA-AAV were then successfully treated with the method. The reanalysis of the samples incurred was below 14 %. This method is suitable for plasma drug monitoring of avacopan because it is both accurate and precise, and meets all validation criteria.

Authors

Institutions

Publication Details

Journal
Journal of Pharmaceutical and Biomedical Analysis
Published
2025-09-22
DOI
https://doi.org/10.1016/j.jpba.2025.117151
Primary Topic
Vasculitis and related conditions
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

LC-MS/MS method for quantification of avacopan in human plasma from patients treated for antineutrophil cytoplasmic antibody-associated vasculitis

J.P. Coindre, Benoı̂t Blanchet, Michel Vidal, Stanislas Faguer et al.
Journal of Pharmaceutical and Biomedical Analysis
Vasculitis and related conditions
article

LC-MS/MS method for quantification of avacopan in human plasma from patients treated for antineutrophil cytoplasmic antibody-associated vasculitis

J.P. Coindre, Benoı̂t Blanchet, Michel Vidal, Stanislas Faguer, Juliette Blondel, Xavier Declèves, Léo Froelicher Bournaud, Benjamin Terrier, Alicja Puzskiel
article en

Abstract

Avacopan is a new treatment for antineutrophil cytoplasmic antibody-associated vasculitis (ANCA-AAV). To date, there has been a lack of published bioanalytical methods to assay it in plasma, resulting in sparse pharmacokinetic data in clinical settings. The objective of this study was to develop and validate a liquid chromatography tandem mass spectrometry (LC-MS/MS) method for the quantification of avacopan in human plasma. [<sup>2</sup>H<sub>4</sub>]-Avacopan was as used as internal standard (IS). Samples were prepared by protein precipitation and separated was on an Accurore® C18 column (2.1 ×50 mm, 2.6 µm), with an elution gradient at a flow rate of 0.5 mL/min. The mobile phase consisted of acetonitrile (0.1 % formic acid) and water (0.1 % formic acid). The analysis run time was 6 min. Avacopan was detected by electrospray ionization on a TSQ Quantis® triple quadrupole mass spectrometer (ThermoFisher Scientific). The linearity of method ranged from 10 to 800 ng/mL. The within-run and between-run relative standard deviations were < 10.2 %. The within-run and between-run relative errors ranged from 2.4 % to 14.4 %. The IS-normalized matrix effect ranged from 2.2 % to 5.1 %, and the IS-normalized extraction recovery ranged from 104.3 % to 109.7 %. The method was fully validated, including checks on linearity, dilution integrity and carry-over. Plasma samples from 16 patients undergoing treatment for ANCA-AAV were then successfully treated with the method. The reanalysis of the samples incurred was below 14 %. This method is suitable for plasma drug monitoring of avacopan because it is both accurate and precise, and meets all validation criteria.

Journal of Pharmaceutical and Biomedical AnalysisVol. 267
Centre National de la Recherche Scientifique (FR), Inserm (FR), Université Paris Cité (FR), Centre Hospitalier Universitaire de Toulouse (FR), Hôpital Cochin (FR), Assistance Publique – Hôpitaux de Paris (FR), Centre Hospitalier du Mans (FR), Metabolism and Renal Physiology (FR)
Openalex Percentile: Top 63%
Vasculitis and related conditions
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.