Novel Blood Collection Tubes Improve Sample Preservation in a Multicenter Study in Thailand

<b>Background</b>: Blood collection tubes (BCTs) are critical in vitro diagnostic devices used in clinical laboratory testing. Innomed tubes are novel BCTs coated with heparin and anti-glycolytic agents (Innomed 1) and clot activators combined with anti-glycolytic agents (Innomed 2). This study, we focus on the verification and multicenter validation of Innomed tubes, aiming to assess their performance in glucose stability, hemolysis resistance, and biochemical interferences relevant to diabetes and non-communicable disease (NCD) monitoring. <b>Methods</b>: Two types of Innomed tubes were evaluated. The verification process, conducted in a reference laboratory, assessed biochemical interferences, glucose stability, and the potential for hemolysis due to centrifugation and blood collection techniques. The multicenter validation of biochemical interferences was carried out across five hospital laboratories. Subsequently, samples were centrifuged at 3500 rpm for 5-10 min and analyzed immediately after separation, not exceeding 2 h from collection. <b>Results</b>: Glucose levels in Innomed 1 and 2 remained stable within 8% of up to 8 h post-collection. No hemolysis was observed under varying centrifugation times (at 3500 rpm) and blood collection techniques, as confirmed through visual inspection and lactate dehydrogenase level determinations. Innomed tubes were suitable BCTs for glucose, HbA1c, thyroid stimulating hormone, free triiodothyronine, free thyroxin, triiodothyronine, carcinoembryonic antigen, and prostate-specific antigen, as well as for 19 routine biochemical assays. <b>Conclusions</b>: Innomed 1 and 2 BCTs maintain blood glucose stability for 8 h, ensuring accurate biochemical, HbA1c, thyroid, and tumor marker testing. Their hemolysis resistance supports diabetes screening and Non-Communicable Diseases (NCD) monitoring. thereby emphasizing their clinical relevance in chronic disease management.

Authors

Institutions

Publication Details

Journal
Diagnostics
Published
2025-09-20
DOI
https://doi.org/10.3390/diagnostics15182398
Primary Topic
Blood groups and transfusion
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Novel Blood Collection Tubes Improve Sample Preservation in a Multicenter Study in Thailand

Khundaw Moonla, Wanvisa Treebuphachatsakul, Napaporn Apiratmateekul, Nam K. Tran et al.
Diagnostics
Blood groups and transfusion
article

Novel Blood Collection Tubes Improve Sample Preservation in a Multicenter Study in Thailand

Khundaw Moonla, Wanvisa Treebuphachatsakul, Napaporn Apiratmateekul, Nam K. Tran, Renu Wiriyaprasit
article en

Abstract

<b>Background</b>: Blood collection tubes (BCTs) are critical in vitro diagnostic devices used in clinical laboratory testing. Innomed tubes are novel BCTs coated with heparin and anti-glycolytic agents (Innomed 1) and clot activators combined with anti-glycolytic agents (Innomed 2). This study, we focus on the verification and multicenter validation of Innomed tubes, aiming to assess their performance in glucose stability, hemolysis resistance, and biochemical interferences relevant to diabetes and non-communicable disease (NCD) monitoring. <b>Methods</b>: Two types of Innomed tubes were evaluated. The verification process, conducted in a reference laboratory, assessed biochemical interferences, glucose stability, and the potential for hemolysis due to centrifugation and blood collection techniques. The multicenter validation of biochemical interferences was carried out across five hospital laboratories. Subsequently, samples were centrifuged at 3500 rpm for 5-10 min and analyzed immediately after separation, not exceeding 2 h from collection. <b>Results</b>: Glucose levels in Innomed 1 and 2 remained stable within 8% of up to 8 h post-collection. No hemolysis was observed under varying centrifugation times (at 3500 rpm) and blood collection techniques, as confirmed through visual inspection and lactate dehydrogenase level determinations. Innomed tubes were suitable BCTs for glucose, HbA1c, thyroid stimulating hormone, free triiodothyronine, free thyroxin, triiodothyronine, carcinoembryonic antigen, and prostate-specific antigen, as well as for 19 routine biochemical assays. <b>Conclusions</b>: Innomed 1 and 2 BCTs maintain blood glucose stability for 8 h, ensuring accurate biochemical, HbA1c, thyroid, and tumor marker testing. Their hemolysis resistance supports diabetes screening and Non-Communicable Diseases (NCD) monitoring. thereby emphasizing their clinical relevance in chronic disease management.

DiagnosticsVol. 15(18)
University of Pittsburgh (US), Computational Diagnostics (United States) (US), Naresuan University (TH)
Openalex Percentile: Top 66%
Blood groups and transfusion
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.