Design and Assessment of Flexible Capacitive Electrodes for Reusable ECG Monitoring: Effects of Sweat and Adapted Front-End Configuration

This work presents the development and characterization of a flexible capacitive electrode for non-contact ECG acquisition, fabricated using a simple and cost-effective method from readily available materials. The electrode consists of a multilayer structure with a copper conductor laminated by a polyimide (Kapton<sup>®</sup>) dielectric layer on a polyurethane support. The impedance and capacitance of the electrode were evaluated under varying textile moisture levels with artificial sweat, as well as after exposure to common disinfectants including ethyl alcohol and iodine tincture. Electrochemical impedance spectroscopy (EIS) and broadband impedance measurements (10<sup>-1</sup>-10<sup>5</sup> Hz) confirmed stable capacitive behavior, moderate sensitivity to moisture, and chemical stability of the Kapton-copper interface under conditions simulating repeated use. A custom front-end readout circuit was implemented to demonstrate through-textile ECG signal acquisition. Simulator tests reproduced characteristic waveform patterns, and preliminary volunteer recordings confirmed the feasibility of through-textile acquisition. These results highlight the promise of the electrode as a low-cost platform for future wearable biosignal monitoring technical research.

Authors

Institutions

Publication Details

Journal
Sensors
Published
2025-09-19
DOI
https://doi.org/10.3390/s25185856
Primary Topic
Advanced Sensor and Energy Harvesting Materials
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Design and Assessment of Flexible Capacitive Electrodes for Reusable ECG Monitoring: Effects of Sweat and Adapted Front-End Configuration

Boriana Tzaneva, Georgi Nikolov, B. Stefanov, Nikolay Tomchev et al.
Sensors
Advanced Sensor and Energy Harvesting Materials
article

Design and Assessment of Flexible Capacitive Electrodes for Reusable ECG Monitoring: Effects of Sweat and Adapted Front-End Configuration

Boriana Tzaneva, Georgi Nikolov, B. Stefanov, Nikolay Tomchev, Ivo Iliev
article en

Abstract

This work presents the development and characterization of a flexible capacitive electrode for non-contact ECG acquisition, fabricated using a simple and cost-effective method from readily available materials. The electrode consists of a multilayer structure with a copper conductor laminated by a polyimide (Kapton<sup>®</sup>) dielectric layer on a polyurethane support. The impedance and capacitance of the electrode were evaluated under varying textile moisture levels with artificial sweat, as well as after exposure to common disinfectants including ethyl alcohol and iodine tincture. Electrochemical impedance spectroscopy (EIS) and broadband impedance measurements (10<sup>-1</sup>-10<sup>5</sup> Hz) confirmed stable capacitive behavior, moderate sensitivity to moisture, and chemical stability of the Kapton-copper interface under conditions simulating repeated use. A custom front-end readout circuit was implemented to demonstrate through-textile ECG signal acquisition. Simulator tests reproduced characteristic waveform patterns, and preliminary volunteer recordings confirmed the feasibility of through-textile acquisition. These results highlight the promise of the electrode as a low-cost platform for future wearable biosignal monitoring technical research.

SensorsVol. 25(18)
Technical University of Sofia (BG)
Openalex Percentile: Top 75%
Advanced Sensor and Energy Harvesting Materials
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.