Study on Factors Affecting Toric Intraocular Lens Rotation Using Intraoperative OCT—Factors Influencing IOL Deployment and Proximity to Posterior Capsule After Insertion

<b>Background/Objectives</b>: Cataract surgery often reveals preexisting corneal astigmatism, which can be corrected using a toric intraocular lens (T-IOL). However, postoperative T-IOL rotation may compromise correction. We investigated T-IOL rotation, focusing on deployment time and proximity to the posterior capsule (PC), using intraoperative optical coherence tomography (iOCT). <b>Methods</b>: Six different T-IOL models were inserted into acrylic simulated lens capsule models under different tacking durations (5 s, 30 s, and 60 s) and temperature conditions (23 °C, 28 °C, and 32 °C). The selection criteria for porcine lenses for examination required that they match human lens dimensions, typical of those used to train cataract surgeons. T-IOL misalignment due to vibration was assessed. Additionally, the impact of temporary intraocular pressure (IOP) reduction on T-IOL proximity to the PC was measured using iOCT in porcine eyes. <b>Results</b>: Tacking time and temperature independently affected T-IOL deployment, with shorter tacking durations and higher temperatures leading to faster deployment. Among lenses tested under identical tacking time and temperature conditions, iSert Micro Toric Aspheric 1-Piece IOL (355T3) had the slowest expansion time, while Avansee™ Preload 1-Piece Toric (YP-T3) had the fastest. Porcine eyes with a corneal white-to-white major axis < 16.0 mm fell within the 95% confidence interval for matching human lens size. Temporarily reducing IOP during surgery improved T-IOL adhesion to the PC, reducing both the occurrence and degree (from 14.0° to nearly 0°) of postoperative rotation. <b>Conclusions</b>: Optimal T-IOL deployment, temporary IOP reduction during surgery, and enhanced adhesion to the PC can reduce the risk and degree of T-IOL rotation. Intraoperative iOCT aids in monitoring T-IOL positioning, which is essential to prevent rotation. Accumulated fluid between the T-IOL and PC may contribute to rotation, which requires further investigation. These findings provide practical strategies for enhancing T-IOL stability and improving the effectiveness of astigmatism correction in cataract surgery.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Medicine
Published
2025-09-19
DOI
https://doi.org/10.3390/jcm14186599
Primary Topic
Intraocular Surgery and Lenses
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Study on Factors Affecting Toric Intraocular Lens Rotation Using Intraoperative OCT—Factors Influencing IOL Deployment and Proximity to Posterior Capsule After Insertion

Hiroto Toda, Naoki Yamamoto, Kei Ichikawa, Kazuo Ichikawa et al.
Journal of Clinical Medicine
Intraocular Surgery and Lenses
article

Study on Factors Affecting Toric Intraocular Lens Rotation Using Intraoperative OCT—Factors Influencing IOL Deployment and Proximity to Posterior Capsule After Insertion

Hiroto Toda, Naoki Yamamoto, Kei Ichikawa, Kazuo Ichikawa, Seiji Tokiwa, Yukihiro Sakai, Yukihito Kato, Yoshiki Tanaka
article en

Abstract

<b>Background/Objectives</b>: Cataract surgery often reveals preexisting corneal astigmatism, which can be corrected using a toric intraocular lens (T-IOL). However, postoperative T-IOL rotation may compromise correction. We investigated T-IOL rotation, focusing on deployment time and proximity to the posterior capsule (PC), using intraoperative optical coherence tomography (iOCT). <b>Methods</b>: Six different T-IOL models were inserted into acrylic simulated lens capsule models under different tacking durations (5 s, 30 s, and 60 s) and temperature conditions (23 °C, 28 °C, and 32 °C). The selection criteria for porcine lenses for examination required that they match human lens dimensions, typical of those used to train cataract surgeons. T-IOL misalignment due to vibration was assessed. Additionally, the impact of temporary intraocular pressure (IOP) reduction on T-IOL proximity to the PC was measured using iOCT in porcine eyes. <b>Results</b>: Tacking time and temperature independently affected T-IOL deployment, with shorter tacking durations and higher temperatures leading to faster deployment. Among lenses tested under identical tacking time and temperature conditions, iSert Micro Toric Aspheric 1-Piece IOL (355T3) had the slowest expansion time, while Avansee™ Preload 1-Piece Toric (YP-T3) had the fastest. Porcine eyes with a corneal white-to-white major axis < 16.0 mm fell within the 95% confidence interval for matching human lens size. Temporarily reducing IOP during surgery improved T-IOL adhesion to the PC, reducing both the occurrence and degree (from 14.0° to nearly 0°) of postoperative rotation. <b>Conclusions</b>: Optimal T-IOL deployment, temporary IOP reduction during surgery, and enhanced adhesion to the PC can reduce the risk and degree of T-IOL rotation. Intraoperative iOCT aids in monitoring T-IOL positioning, which is essential to prevent rotation. Accumulated fluid between the T-IOL and PC may contribute to rotation, which requires further investigation. These findings provide practical strategies for enhancing T-IOL stability and improving the effectiveness of astigmatism correction in cataract surgery.

Journal of Clinical MedicineVol. 14(18)
Fujita Health University (JP), Chukyo Hospital (JP)
Openalex Percentile: Top 62%
Intraocular Surgery and Lenses
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.