Development of a Novel Algorithm for Tip Fold-Over Detection in Cochlear Implants and Evaluation on Bench and Multiple Clinical Data Bases

<b>Objectives:</b> Tip fold-over (TFO) is a rare but critical occurrence in cochlear implant procedures where the electrode array folds back on itself within the cochlea, compromising programming and device performance. Timely intraoperative detection is essential for immediate correction and optimal placement. Electric field imaging (EFI) has shown promise for identifying TFO both intra- and post-operatively. This study evaluates the performance of a TFO detection algorithm implemented in Target CI (Version 1.6) using Advanced Bionics' cochlear implant systems, validated through bench and patient datasets. <b>Methods:</b> Sample data included (1) bench testing with a plastic cochlea and human temporal bones with and without induced TFOs, confirmed visually or radiographically; (2) intraoperative EFI measurements recorded using the AIM™ system, with electrode placement confirmed through imaging; and (3) historical EFI recordings from the Target CI DataLake, which lacks imaging and programming metadata. The TFO algorithm's performance was evaluated by assessing its sensitivity and specificity using these datasets. <b>Results:</b> The TFO algorithm achieved 100% sensitivity and specificity in bench models and intraoperative EFI with imaging-confirmed placements. Among 226 intra-op cases, four TFOs were confirmed by imaging, and all were correctly identified by the algorithm. In the large set of DataLake cases (14,734 implants), 0.80% were flagged as potential TFOs. TFO prevalence was higher with pre-curved arrays (1.22%) than straight lateral wall arrays (0.32%). <b>Conclusions:</b> The TFO algorithm showed high reliability with 100% sensitivity and specificity using routine clinical EFI data. While not a replacement for imaging, the TFO algorithm serves as a fast, accessible tool to alert clinicians to potential TFOs.

Authors

Institutions

Publication Details

Journal
Audiology Research
Published
2025-09-18
DOI
https://doi.org/10.3390/audiolres15050118
Citations
1
Primary Topic
Hearing Loss and Rehabilitation
Type
article
Field-Weighted Citation Impact
1.43
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development of a Novel Algorithm for Tip Fold-Over Detection in Cochlear Implants and Evaluation on Bench and Multiple Clinical Data Bases

Ashish Mehta, Tony Spahr, Jordan Varghese, Syed F. Ahsan et al.
1 citations
Audiology Research
Hearing Loss and Rehabilitation
1.43
article

Development of a Novel Algorithm for Tip Fold-Over Detection in Cochlear Implants and Evaluation on Bench and Multiple Clinical Data Bases

Ashish Mehta, Tony Spahr, Jordan Varghese, Syed F. Ahsan, Christopher Danner, Andrea J. DeFreese, Kanthaiah Koka, Kyle P. Allen, Craig A. Buchman, Loren Bartels, Matthew Miller, Patrick Boyle, Mehrangiz Ashiri, Azret Botash
article en
1 citations

Abstract

<b>Objectives:</b> Tip fold-over (TFO) is a rare but critical occurrence in cochlear implant procedures where the electrode array folds back on itself within the cochlea, compromising programming and device performance. Timely intraoperative detection is essential for immediate correction and optimal placement. Electric field imaging (EFI) has shown promise for identifying TFO both intra- and post-operatively. This study evaluates the performance of a TFO detection algorithm implemented in Target CI (Version 1.6) using Advanced Bionics' cochlear implant systems, validated through bench and patient datasets. <b>Methods:</b> Sample data included (1) bench testing with a plastic cochlea and human temporal bones with and without induced TFOs, confirmed visually or radiographically; (2) intraoperative EFI measurements recorded using the AIM™ system, with electrode placement confirmed through imaging; and (3) historical EFI recordings from the Target CI DataLake, which lacks imaging and programming metadata. The TFO algorithm's performance was evaluated by assessing its sensitivity and specificity using these datasets. <b>Results:</b> The TFO algorithm achieved 100% sensitivity and specificity in bench models and intraoperative EFI with imaging-confirmed placements. Among 226 intra-op cases, four TFOs were confirmed by imaging, and all were correctly identified by the algorithm. In the large set of DataLake cases (14,734 implants), 0.80% were flagged as potential TFOs. TFO prevalence was higher with pre-curved arrays (1.22%) than straight lateral wall arrays (0.32%). <b>Conclusions:</b> The TFO algorithm showed high reliability with 100% sensitivity and specificity using routine clinical EFI data. While not a replacement for imaging, the TFO algorithm serves as a fast, accessible tool to alert clinicians to potential TFOs.

Audiology ResearchVol. 15(5)
Sanford Health (US), Washington University in St. Louis (US), Advanced Technology and Research Corporation (United States) (US), Kaiser Permanente Anaheim Medical Center (US), Vanderbilt University Medical Center (US), University of Tampa (US)
Openalex Percentile: Top 15%
Hearing Loss and Rehabilitation
1.43
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.