Development of long working distance reflection optics for soft x-ray reflection ptychography

Soft x-ray microscopy is increasingly applied in fields such as biology and solid-state physics. Among soft x-ray imaging techniques, ptychography is a method that retrieves phase information from coherent diffraction patterns with iterative calculations. Ptychography enables high spatial resolution and field of view independent of the optical setup. However, conventional soft x-ray ptychography experiments require the sample to be thin to ensure sufficient transmission. To overcome this limitation, we propose a new reflective optical setup for x-ray ptychography, where the sample is tilted significantly to reflect soft x-rays from its surface. This configuration allows for the analysis of thick and opaque samples, greatly expanding the range of materials and conditions that can be investigated. Furthermore, it will enable surface imaging, which is difficult for the standard transmission geometry.

Authors

Institutions

Publication Details

Published
2025-09-18
DOI
https://doi.org/10.1117/12.3063334
Primary Topic
Advanced X-ray Imaging Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development of long working distance reflection optics for soft x-ray reflection ptychography

Takenori Shimamura, Haruhiko Ohashi, Yoko Takeo, Kyota Yoshinaga et al.
Advanced X-ray Imaging Techniques
article

Development of long working distance reflection optics for soft x-ray reflection ptychography

Takenori Shimamura, Haruhiko Ohashi, Yoko Takeo, Kyota Yoshinaga, Kai Sakurai, Yasunori Senba, Makina Yabashi, Takashi Kimura, Jordan T. O’Neal, Hikaru Kishimoto, Yu Nakata, Y. Nagayama
article en

Abstract

Soft x-ray microscopy is increasingly applied in fields such as biology and solid-state physics. Among soft x-ray imaging techniques, ptychography is a method that retrieves phase information from coherent diffraction patterns with iterative calculations. Ptychography enables high spatial resolution and field of view independent of the optical setup. However, conventional soft x-ray ptychography experiments require the sample to be thin to ensure sufficient transmission. To overcome this limitation, we propose a new reflective optical setup for x-ray ptychography, where the sample is tilted significantly to reflect soft x-rays from its surface. This configuration allows for the analysis of thick and opaque samples, greatly expanding the range of materials and conditions that can be investigated. Furthermore, it will enable surface imaging, which is difficult for the standard transmission geometry.

SPring-8 (JP), Japan Synchrotron Radiation Research Institute (JP), The University of Tokyo (JP)
Openalex Percentile: Top 68%
Advanced X-ray Imaging Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.