Long-Term Evaluation of Bone Healing Monitoring Using an Instrumented Plate with Measurement Sensors (Smart Implant) over 10 Years

A total of 66 smart implants were included. As a measure of bony stability, the relative elastic compliance of the osteosynthesis was determined from the gradient between the applied external load and the measured implant load over the entire healing process. The healing process of non-unions of the femur with a smart implant was tracked by telemetric measurements over a timespan of up to 10 years. The measurements of the longest healing process show a very slow but constant decrease in force transmission over the implant, radiological findings over 10 years show corresponding consolidation until bony healing. The use of a telemetrically instrumented bone plate, a so-called smart implant, to monitor the healing process is a successful procedure to support the clinician in his decision to take further surgical measures or to wait until healing occurs.

Authors

Institutions

Publication Details

Journal
Sensors
Published
2025-09-16
DOI
https://doi.org/10.3390/s25185779
Citations
2
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
4.01
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Long-Term Evaluation of Bone Healing Monitoring Using an Instrumented Plate with Measurement Sensors (Smart Implant) over 10 Years

Arndt‐Peter Schulz, B. Kienast, Tobias Barth, Birgitt Kowald et al.
2 citations
Sensors
Orthopaedic implants and arthroplasty
4.01
article

Long-Term Evaluation of Bone Healing Monitoring Using an Instrumented Plate with Measurement Sensors (Smart Implant) over 10 Years

Arndt‐Peter Schulz, B. Kienast, Tobias Barth, Birgitt Kowald, N. Weinrich, Matthias Münch, Klaus Seide
article en
2 citations

Abstract

A total of 66 smart implants were included. As a measure of bony stability, the relative elastic compliance of the osteosynthesis was determined from the gradient between the applied external load and the measured implant load over the entire healing process. The healing process of non-unions of the femur with a smart implant was tracked by telemetric measurements over a timespan of up to 10 years. The measurements of the longest healing process show a very slow but constant decrease in force transmission over the implant, radiological findings over 10 years show corresponding consolidation until bony healing. The use of a telemetrically instrumented bone plate, a so-called smart implant, to monitor the healing process is a successful procedure to support the clinician in his decision to take further surgical measures or to wait until healing occurs.

SensorsVol. 25(18)
Fraunhofer-Einrichtung für Individualisierte Medizintechnik (DE), BG Klinikum Hamburg (DE), University of Lübeck (DE)
Openalex Percentile: Top 6%
Orthopaedic implants and arthroplasty
4.01
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.