The Influence of Reaming Velocity During Preparation of the Femoral Canal—An In Vitro Analysis of Two Straight Femoral Revision Stems with a Fluted Tapered Design

<i>Background:</i> The use of tapered fluted revision stems has been shown to be reliable and safe. Primary stability is mandatory for a long-lasting fixation between bone and a prosthesis. Nevertheless, aseptic loosening due to insufficient primary stability occurs and may be related to technically improper preparation of the femoral canal. Instructions of manufacturers are heterogeneous regarding preparation of implant beds. <i>Questions/Purposes:</i> Does speed or the design of the reamer influence the accuracy of the implant bed and, consecutively, primary stability? <i>Materials and Methods:</i> A test foam with an elastic moduli and pressure resistance similar to that of cancellous bone was used. The medullary canal was prepared with the use of reamers of two different straight and tapered femoral revision devices. Three different rotational speeds were used for preparation. After preparation, primary stability was measured and fixating characteristics were derived. <i>Results:</i> Sufficient primary stability was achievable by all three preparation methods but fixating characteristics were different. Significantly higher micro-motions were detected near the tip of the prosthesis compared to those at all more proximal measuring points. Reaming with high velocity resulted in significantly higher micro-motions compared to that with mid- or low-speed burring. <i>Conclusions:</i> Different preparation methods may be one explanation for the range of reported survivorship data of the two devices with aseptic loosening as the end point. The precision of the implant bed and fixating characteristics were best after reaming with lower velocity. Superior but not significantly better fixation characteristics were achieved with the monobloc stem compared to those with the modular device.

Authors

Institutions

Publication Details

Journal
Bioengineering
Published
2025-09-16
DOI
https://doi.org/10.3390/bioengineering12090984
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The Influence of Reaming Velocity During Preparation of the Femoral Canal—An In Vitro Analysis of Two Straight Femoral Revision Stems with a Fluted Tapered Design

Matthias K. Jung, Eike Jakubowitz, Arnold J. Suda, Oliver E. Bischel et al.
Bioengineering
Orthopaedic implants and arthroplasty
article

The Influence of Reaming Velocity During Preparation of the Femoral Canal—An In Vitro Analysis of Two Straight Femoral Revision Stems with a Fluted Tapered Design

Matthias K. Jung, Eike Jakubowitz, Arnold J. Suda, Oliver E. Bischel, Jörn Bengt Seeger, Alexander Jahnke, Stefan Dörfler
article en

Abstract

<i>Background:</i> The use of tapered fluted revision stems has been shown to be reliable and safe. Primary stability is mandatory for a long-lasting fixation between bone and a prosthesis. Nevertheless, aseptic loosening due to insufficient primary stability occurs and may be related to technically improper preparation of the femoral canal. Instructions of manufacturers are heterogeneous regarding preparation of implant beds. <i>Questions/Purposes:</i> Does speed or the design of the reamer influence the accuracy of the implant bed and, consecutively, primary stability? <i>Materials and Methods:</i> A test foam with an elastic moduli and pressure resistance similar to that of cancellous bone was used. The medullary canal was prepared with the use of reamers of two different straight and tapered femoral revision devices. Three different rotational speeds were used for preparation. After preparation, primary stability was measured and fixating characteristics were derived. <i>Results:</i> Sufficient primary stability was achievable by all three preparation methods but fixating characteristics were different. Significantly higher micro-motions were detected near the tip of the prosthesis compared to those at all more proximal measuring points. Reaming with high velocity resulted in significantly higher micro-motions compared to that with mid- or low-speed burring. <i>Conclusions:</i> Different preparation methods may be one explanation for the range of reported survivorship data of the two devices with aseptic loosening as the end point. The precision of the implant bed and fixating characteristics were best after reaming with lower velocity. Superior but not significantly better fixation characteristics were achieved with the monobloc stem compared to those with the modular device.

BioengineeringVol. 12(9)
Justus-Liebig-Universität Gießen (DE), Heidelberg University (DE), University Hospital Heidelberg (DE), Medizinische Hochschule Hannover (DE), Ansbach University of Applied Sciences (DE), Institut Kurz (DE), University Medical Centre Mannheim (DE)
Openalex Percentile: Top 60%
Orthopaedic implants and arthroplasty
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.