Ocular Biometry and Refractive Prediction in Short Eyes: A Comparison of Two Swept-Source Optical Coherence Tomography-Based Biometers

<b><i>Purpose:</i></b> To compare the performance of two swept-source optical coherence tomography-based biometers in the measurement of ocular biometry and the prediction of postoperative refractive errors in eyes with short axial length (AL). <b><i>Methods:</i></b> A total of 48 eyes from 29 patients with AL ≤ 22 mm were included. AL, anterior chamber depth (ACD), keratometry (K), and lens thickness (LT) measured using the IOLMaster<sup>®</sup> 700 and ARGOS<sup>®</sup> before cataract surgery were compared. The refractive error prediction accuracy of the two devices was also compared. <b><i>Results:</i></b> This study included four men (7 eyes) and 25 women (41 eyes), with an average age of 70.7 ± 8.1 years (mean ± SD; range, 47-82 years). The two devices demonstrated good agreement in measurements of ocular biometry with high intraclass correlation coefficients (AL = 0.975; ACD = 0.957; K = 0.988; LT = 0.994). However, AL and ACD were significantly shorter when measured with the IOLMaster<sup>®</sup> 700 compared to the ARGOS<sup>®</sup> (<i>p</i> < 0.001 for both). There was no significant difference in mean absolute prediction errors between the two devices (<i>p</i> = 0.423). The IOLMaster<sup>®</sup> 700 showed a significantly lower mean prediction error than the ARGOS<sup>®</sup> (+0.12 ± 0.39 diopters vs. +0.20 ± 0.39 diopters, <i>p</i> = 0.006), although the difference was of limited clinical relevance. There were no significant differences in the percentages of eyes within ± 0.50 D (77.1% vs. 75.0%, <i>p</i> = 0.811) and ± 1.00 D (100% vs. 97.9%, <i>p</i> = 0.315) of the predicted refractive error. <b><i>Conclusions:</i></b> Although IOLMaster<sup>®</sup> 700 and ARGOS<sup>®</sup> showed good agreements in eyes with short AL, significant differences were observed in the measurements of AL and ACD. Both devices demonstrated good efficacy and comparable performance in predicting postoperative refractive errors.

Authors

Publication Details

Journal
Bioengineering
Published
2025-09-16
DOI
https://doi.org/10.3390/bioengineering12090983
Citations
2
Primary Topic
Glaucoma and retinal disorders
Type
article
Field-Weighted Citation Impact
3.41
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Ocular Biometry and Refractive Prediction in Short Eyes: A Comparison of Two Swept-Source Optical Coherence Tomography-Based Biometers

Sang Beom Han, Ji Su Seong
2 citations
Bioengineering
Glaucoma and retinal disorders
3.41
article

Ocular Biometry and Refractive Prediction in Short Eyes: A Comparison of Two Swept-Source Optical Coherence Tomography-Based Biometers

Sang Beom Han, Ji Su Seong
article en
2 citations

Abstract

<b><i>Purpose:</i></b> To compare the performance of two swept-source optical coherence tomography-based biometers in the measurement of ocular biometry and the prediction of postoperative refractive errors in eyes with short axial length (AL). <b><i>Methods:</i></b> A total of 48 eyes from 29 patients with AL ≤ 22 mm were included. AL, anterior chamber depth (ACD), keratometry (K), and lens thickness (LT) measured using the IOLMaster<sup>®</sup> 700 and ARGOS<sup>®</sup> before cataract surgery were compared. The refractive error prediction accuracy of the two devices was also compared. <b><i>Results:</i></b> This study included four men (7 eyes) and 25 women (41 eyes), with an average age of 70.7 ± 8.1 years (mean ± SD; range, 47-82 years). The two devices demonstrated good agreement in measurements of ocular biometry with high intraclass correlation coefficients (AL = 0.975; ACD = 0.957; K = 0.988; LT = 0.994). However, AL and ACD were significantly shorter when measured with the IOLMaster<sup>®</sup> 700 compared to the ARGOS<sup>®</sup> (<i>p</i> < 0.001 for both). There was no significant difference in mean absolute prediction errors between the two devices (<i>p</i> = 0.423). The IOLMaster<sup>®</sup> 700 showed a significantly lower mean prediction error than the ARGOS<sup>®</sup> (+0.12 ± 0.39 diopters vs. +0.20 ± 0.39 diopters, <i>p</i> = 0.006), although the difference was of limited clinical relevance. There were no significant differences in the percentages of eyes within ± 0.50 D (77.1% vs. 75.0%, <i>p</i> = 0.811) and ± 1.00 D (100% vs. 97.9%, <i>p</i> = 0.315) of the predicted refractive error. <b><i>Conclusions:</i></b> Although IOLMaster<sup>®</sup> 700 and ARGOS<sup>®</sup> showed good agreements in eyes with short AL, significant differences were observed in the measurements of AL and ACD. Both devices demonstrated good efficacy and comparable performance in predicting postoperative refractive errors.

BioengineeringVol. 12(9)
Openalex Percentile: Top 7%
Glaucoma and retinal disorders
3.41
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Ocular Biometry and Refractive Prediction in Short Eyes: A Comparison of Two Swept-Source Optical Coherence Tomography-Based Biometers — Sang Beom Han, Ji Su Seong · Bioengineering (2025) | TGRS Research Map | TGRS