The article presents a method to stabilise the output pulse for the SiPM scintillation detector, which uses CLYC(Ce) scintillation crystal coupled with a silicon photomultiplier (SiPM) to measure gamma and neutron radiation.This method is based on the characteristic of SiPM that its gain increases with increasing bias voltage and decreases with increasing temperature.A bias voltage supply was designed for SiPM that uses a microprocessor to manage and control the output voltage value according to the operating ambient temperature.The corrected data on the temperature dependence of the 662 keV peak energy of 137 Cs radioactive source are stored in the memory.Depending on the measured ambient temperature, the microprocessor controls the bias voltage value for SiPM.The results show that in the operating temperature range of the detector from 0.5 to 45 o C, the change of 662 keV peak energy position when uncorrected and corrected is 247 and 5 channels, respectively.Thus, the method of correcting the temperature dependence of the SiPM output pulse amplitude in this study can be widely developed in scintillation detectors using SiPM to replace the photomultiplier tube (PMT).

Authors

Institutions

Publication Details

Journal
Ministry of Science and Technology Vietnam
Published
2025-08-25
DOI
https://doi.org/10.31276/vjst.67(8).07-11
Primary Topic
Radiation Detection and Scintillator Technologies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Correcting the temperature dependence of the SiPM scintillation detector

Thành Hưng Nguyễn, Van Sy Nguyen, Cuong Quoc Dang, Quang Bao Dang
Ministry of Science and Technology Vietnam
Radiation Detection and Scintillator Technologies
article

Correcting the temperature dependence of the SiPM scintillation detector

Thành Hưng Nguyễn, Van Sy Nguyen, Cuong Quoc Dang, Quang Bao Dang
article en

Abstract

The article presents a method to stabilise the output pulse for the SiPM scintillation detector, which uses CLYC(Ce) scintillation crystal coupled with a silicon photomultiplier (SiPM) to measure gamma and neutron radiation.This method is based on the characteristic of SiPM that its gain increases with increasing bias voltage and decreases with increasing temperature.A bias voltage supply was designed for SiPM that uses a microprocessor to manage and control the output voltage value according to the operating ambient temperature.The corrected data on the temperature dependence of the 662 keV peak energy of 137 Cs radioactive source are stored in the memory.Depending on the measured ambient temperature, the microprocessor controls the bias voltage value for SiPM.The results show that in the operating temperature range of the detector from 0.5 to 45 o C, the change of 662 keV peak energy position when uncorrected and corrected is 247 and 5 channels, respectively.Thus, the method of correcting the temperature dependence of the SiPM output pulse amplitude in this study can be widely developed in scintillation detectors using SiPM to replace the photomultiplier tube (PMT).

Ministry of Science and Technology VietnamVol. 67
Institute for Nuclear Science and Technology (VN)
Openalex Percentile: Top 68%
Radiation Detection and Scintillator Technologies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Correcting the temperature dependence of the SiPM scintillation detector — Thành Hưng Nguyễn, Van Sy Nguyen, et al. · Ministry of Science and Technology Vietnam (2025) | TGRS Research Map | TGRS