When Legacy Triumphs: Resolving Leadless Pacemaker Programming Failure Caused by Loop Recorder Interference

The growing prevalence of co-implanted cardiac electronic devices increases the potential for clinically significant interactions. We describe the case of an 81-year-old man with a Biotronik BIOMONITOR IV implantable loop recorder (ILR) who required a Medtronic MICRA VR leadless pacemaker. Post-implantation, interrogation with the modern Medtronic SmartSync programmer failed due to suspected electromagnetic interference from the ILR. Communication was successfully established using the older Medtronic CareLink 2090 programmer. This report highlights a unique instance of inter-device interference, underscoring the importance of understanding programmer technology and device compatibility in complex clinical scenarios.

Authors

Institutions

Publication Details

Journal
Pacing and Clinical Electrophysiology
Published
2025-09-17
DOI
https://doi.org/10.1111/pace.70055
Citations
1
Primary Topic
Healthcare Technology and Patient Monitoring
Type
article
Field-Weighted Citation Impact
2.01
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

When Legacy Triumphs: Resolving Leadless Pacemaker Programming Failure Caused by Loop Recorder Interference

Elisabetta Fratini, Antonino Previti, Gianpaolo Varalda, Paolo Garrone et al.
1 citations
Pacing and Clinical Electrophysiology
Healthcare Technology and Patient Monitoring
2.01
article

When Legacy Triumphs: Resolving Leadless Pacemaker Programming Failure Caused by Loop Recorder Interference

Elisabetta Fratini, Antonino Previti, Gianpaolo Varalda, Paolo Garrone, Alessandra Chinaglia, Alfonso Franzè, G. Pavani
article en
1 citations

Abstract

The growing prevalence of co-implanted cardiac electronic devices increases the potential for clinically significant interactions. We describe the case of an 81-year-old man with a Biotronik BIOMONITOR IV implantable loop recorder (ILR) who required a Medtronic MICRA VR leadless pacemaker. Post-implantation, interrogation with the modern Medtronic SmartSync programmer failed due to suspected electromagnetic interference from the ILR. Communication was successfully established using the older Medtronic CareLink 2090 programmer. This report highlights a unique instance of inter-device interference, underscoring the importance of understanding programmer technology and device compatibility in complex clinical scenarios.

Pacing and Clinical ElectrophysiologyVol. 48(11)
Ospedale San Luigi Gonzaga (IT), University of Turin (IT)
Openalex Percentile: Top 12%
Healthcare Technology and Patient Monitoring
2.01
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

When Legacy Triumphs: Resolving Leadless Pacemaker Programming Failure Caused by Loop Recorder Interference — Elisabetta Fratini, Antonino Previti, et al. · Pacing and Clinical Electrophysiology (2025) | TGRS Research Map | TGRS