A compact stimulator and recorder for atrial fibrillation ablation surgery: Realization and clinical evaluation

This paper details the realization and testing of a new compact instrument optimized for use during cardiac ablation surgery. In contrast to previous research-grade instrumentation, the proposed instrument supports both recording and stimulation operations. It is characterized by an input-referred noise level of 3 μ V RMS and is capable of recording up to 48 cardiac catheter channels at a sampling rate of 1–4 KHz with a resolution of 24-bits and a maximum analogue bandwidth of 0.1–1000 Hz. Two bipolar stimulation signals can be generated in either current-mode or voltage-mode with amplitude of up to 80 mA at 14 V. Additionally, hardware blanking has been implemented to minimize the impact of stimulation artifacts on recordings. Custom firmware and software to control the instrument using a PC have been developed in conjunction with a specialist medical team. The device has been characterized electronically and tested on porcine Lagendorff heart preparations as well as in a typical clinical setup during a cardiac intervention procedure. The proposed device occupies a portable footprint of 19×14×7 cm and can replace two instruments found in a typical electrophysiology lab, a dedicated stimulator and a data acquisition unit, simplifying the hardware and software setup and thus streamlining the testing of new approaches in cardiac treatment. • New compact stimulation and recording instrument for cardiac electrophysiology. • The device enables clinicians to perform both the complex ganglion-plexi (GP) node-based atrial fibrillation ablation procedure and the pulmonary vein isolation (PVI) procedure. • Research is a collaboration between engineers and interventional cardiologists. • Electronic design, realisation, dry lab-based characterisation and wet lab-based evaluation. • Device use demonstrated on-humans highlighting its effectiveness as a clinical tool.

Authors

Institutions

Publication Details

Journal
Measurement
Published
2025-09-16
DOI
https://doi.org/10.1016/j.measurement.2025.118777
Primary Topic
Atrial Fibrillation Management and Outcomes
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A compact stimulator and recorder for atrial fibrillation ablation surgery: Realization and clinical evaluation

Joe Brook, Jamie Kay, Prapa Kanagaratnam, Nick Linton et al.
Measurement
Atrial Fibrillation Management and Outcomes
article

A compact stimulator and recorder for atrial fibrillation ablation surgery: Realization and clinical evaluation

Joe Brook, Jamie Kay, Prapa Kanagaratnam, Nick Linton, Clare Coyle, Rasheda A. Chowdhury, Simos Koutsoftidis, Konstantinos Petkos, Shengzhe Li, Emmanuel M. Drakakis, Min-young Kim, Georgios C. Zafeiropoulos, Burak Temelkuran, Sally A. Bissett, Yacine Belgaid, Georgios Gryparis, Oskar Bushrod
article en

Abstract

This paper details the realization and testing of a new compact instrument optimized for use during cardiac ablation surgery. In contrast to previous research-grade instrumentation, the proposed instrument supports both recording and stimulation operations. It is characterized by an input-referred noise level of 3 μ V RMS and is capable of recording up to 48 cardiac catheter channels at a sampling rate of 1–4 KHz with a resolution of 24-bits and a maximum analogue bandwidth of 0.1–1000 Hz. Two bipolar stimulation signals can be generated in either current-mode or voltage-mode with amplitude of up to 80 mA at 14 V. Additionally, hardware blanking has been implemented to minimize the impact of stimulation artifacts on recordings. Custom firmware and software to control the instrument using a PC have been developed in conjunction with a specialist medical team. The device has been characterized electronically and tested on porcine Lagendorff heart preparations as well as in a typical clinical setup during a cardiac intervention procedure. The proposed device occupies a portable footprint of 19×14×7 cm and can replace two instruments found in a typical electrophysiology lab, a dedicated stimulator and a data acquisition unit, simplifying the hardware and software setup and thus streamlining the testing of new approaches in cardiac treatment. • New compact stimulation and recording instrument for cardiac electrophysiology. • The device enables clinicians to perform both the complex ganglion-plexi (GP) node-based atrial fibrillation ablation procedure and the pulmonary vein isolation (PVI) procedure. • Research is a collaboration between engineers and interventional cardiologists. • Electronic design, realisation, dry lab-based characterisation and wet lab-based evaluation. • Device use demonstrated on-humans highlighting its effectiveness as a clinical tool.

MeasurementVol. 257
Imperial College Healthcare NHS Trust (GB), Imperial College London (GB)
Openalex Percentile: Top 61%
Atrial Fibrillation Management and Outcomes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.