Differences in Arterial Oxygen Saturation Measurements Between Two Co-Oximeters From Different Manufacturers

<b>Background:</b> Functional arterial hemoglobin oxygen saturation is a critical clinical parameter and the reference for verifying accuracy of S<sub>pO<sub>2</sub></sub> values during performance testing of a pulse oximeter. Variable performance between blood gas analyzers that perform co-oximetry may complicate interpretation of research studies and regulatory guidance. This study quantifies systematic differences between multiple same model analyzers from two widely used brands of co-oximeters. <b>Methods:</b> Three sequential experiments were conducted using controlled hypoxemia in healthy adult volunteers according to regulatory clearance criteria for use of functional arterial oxygenation from co-oximetry analysis (sO<sub>2</sub>), to simultaneous S<sub>pO<sub>2</sub></sub> values over the range of 70-100%. To account for potential methodological confounders in experiments 1 and 2, a third experiment analyzed 31 arterial samples across multiple same model analyzers for co-oximetry (two Werfen GEM Premier 5000, and three Radiometer ABL90 Flex) in a randomized order after simultaneous preparation. Statistical analysis included paired differences, mixed-effects modeling, and variance analysis to isolate analyzer-specific effects while controlling for confounding variables. <b>Results:</b> The GEM analyzers consistently measured sO<sub>2</sub> 2.3% higher on average than ABL analyzers (95% CI 2.01-2.63%, <i>P</i> < .001). This difference was more pronounced at lower saturations (3.2% in the sO<sub>2</sub> 70s range vs 1.4% in the 90s range). Despite strong correlation between measurements (r = 0.999), 96.8% of paired samples differed by >1% and 61.3% by >2%. Within-manufacturer variability was similar between brands (mean absolute difference: ABL 0.38%, GEM 0.47%, <i>P</i> = .55). <b>Conclusions:</b> Clinically important systematic differences in sO<sub>2</sub> measurement exist between GEM and ABL blood analyzers that perform co-oximetry, with potential implications for patient care and pulse oximeter verification studies. Larger studies with more co-oximeter brands, coupled with blood tonometry, are needed to better characterize these findings. In the meantime, regulatory standards and reports of pulse oximeter clinical trials should account for these findings and specify which manufacturer's co-oximeters were used.

Authors

Institutions

Publication Details

Journal
Respiratory Care
Published
2025-09-16
DOI
https://doi.org/10.1177/19433654251376285
Citations
2
Primary Topic
Chronic Obstructive Pulmonary Disease (COPD) Research
Type
article
Field-Weighted Citation Impact
3.04
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Differences in Arterial Oxygen Saturation Measurements Between Two Co-Oximeters From Different Manufacturers

Robert J. Kopotic, Carolyn M. Hendrickson, Tyler J. Law, Rocío Zamora et al.
2 citations
Respiratory Care
Chronic Obstructive Pulmonary Disease (COPD) Research
3.04
article

Differences in Arterial Oxygen Saturation Measurements Between Two Co-Oximeters From Different Manufacturers

Robert J. Kopotic, Carolyn M. Hendrickson, Tyler J. Law, Rocío Zamora, Michael S. Lipnick, Philip E. Bickler, John Feiner, Xander M.R. van Wijk, Gregory Leeb, Ella Behnke
article en
2 citations

Abstract

<b>Background:</b> Functional arterial hemoglobin oxygen saturation is a critical clinical parameter and the reference for verifying accuracy of S<sub>pO<sub>2</sub></sub> values during performance testing of a pulse oximeter. Variable performance between blood gas analyzers that perform co-oximetry may complicate interpretation of research studies and regulatory guidance. This study quantifies systematic differences between multiple same model analyzers from two widely used brands of co-oximeters. <b>Methods:</b> Three sequential experiments were conducted using controlled hypoxemia in healthy adult volunteers according to regulatory clearance criteria for use of functional arterial oxygenation from co-oximetry analysis (sO<sub>2</sub>), to simultaneous S<sub>pO<sub>2</sub></sub> values over the range of 70-100%. To account for potential methodological confounders in experiments 1 and 2, a third experiment analyzed 31 arterial samples across multiple same model analyzers for co-oximetry (two Werfen GEM Premier 5000, and three Radiometer ABL90 Flex) in a randomized order after simultaneous preparation. Statistical analysis included paired differences, mixed-effects modeling, and variance analysis to isolate analyzer-specific effects while controlling for confounding variables. <b>Results:</b> The GEM analyzers consistently measured sO<sub>2</sub> 2.3% higher on average than ABL analyzers (95% CI 2.01-2.63%, <i>P</i> < .001). This difference was more pronounced at lower saturations (3.2% in the sO<sub>2</sub> 70s range vs 1.4% in the 90s range). Despite strong correlation between measurements (r = 0.999), 96.8% of paired samples differed by >1% and 61.3% by >2%. Within-manufacturer variability was similar between brands (mean absolute difference: ABL 0.38%, GEM 0.47%, <i>P</i> = .55). <b>Conclusions:</b> Clinically important systematic differences in sO<sub>2</sub> measurement exist between GEM and ABL blood analyzers that perform co-oximetry, with potential implications for patient care and pulse oximeter verification studies. Larger studies with more co-oximeter brands, coupled with blood tonometry, are needed to better characterize these findings. In the meantime, regulatory standards and reports of pulse oximeter clinical trials should account for these findings and specify which manufacturer's co-oximeters were used.

Respiratory CareVol. 70(12)
University of California, San Francisco (US), University of California System (US), Kohler (New Zealand) (NZ), The Medical Device (United Kingdom) (GB), Hospital Vargas (VE)
Openalex Percentile: Top 8%
Chronic Obstructive Pulmonary Disease (COPD) Research
3.04
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.