Quality factor measurement and enhancement techniques for ultra-high field surface loop receiver coils
Motivation: There exists a discrepancy between simulated coil array SNR and experimentally obtained SNR which must be reconciled. Goal(s): To develop useful bench measurements and simulations to predict experimental fraction of ultimate intrinsic SNR achievable with a given practical coil construction. Approach: We propose bench measurements and simulation workflow to identify sources of SNR loss and propose practical strategies to mitigate these losses. Results: Bench measurements and simulations match closely and we provide plots of Q-ratios and fraction of ideal SNR for various coil sizes at 7T and 10.5T. Impact: Bench measurements and simulation results of Q-ratio and SNR fraction vs coil size shared herein provide a useful guide to coil builders in understanding potential sources of SNR losses and estimating those losses in ultra-high field RF coil development.
Authors
- Gregor Adriany (ORCID: https://orcid.org/0000-0002-6428-9005)
- Matt Waks (ORCID: https://orcid.org/0000-0003-1743-8888)
- Russell Lagore (ORCID: https://orcid.org/0000-0003-1950-1828)
- Alireza Sadeghi‐Tarakameh (ORCID: https://orcid.org/0000-0001-5718-6553)
- Kâmil Uǧurbil (ORCID: https://orcid.org/0000-0002-8475-9334)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/4990
- Primary Topic
- Particle accelerators and beam dynamics
- Type
- article
- Field-Weighted Citation Impact
- 0.00