Influence of Deactivated 8Tx/16Rx Transceiver Coil Elements on 3D SAR Distribution at 7T MRI of the Body Trunk
Motivation: In 7T cardiac MRI, controlling peak SAR and SAR distribution is critical for safety. This study explores how deactivating loop coil elements affects these metrics. Goal(s): We assess the impact of coil element deactivation on SAR in a shared-conductor loop coil across different human models. Approach: Electromagnetic simulations were conducted on ten human models, analyzing SAR under normal conditions and with individual coil element deactivations. Results: Coil element deactivation increases peak SAR, up to 77.61% in the Rosalind model, and changes SAR distribution by up to 474% in the Laura model, varying with coil position and model anatomy. Impact: This study highlights the significant impact of coil element deactivation on SAR, requiring a safety factor in clinical routines to protect patients if coil elements become deactivated.
Authors
- Farzad Jabbarigargari (ORCID: https://orcid.org/0000-0002-4468-5721)
- Maxim Terekhov (ORCID: https://orcid.org/0000-0001-5617-7751)
- Wolfgang Schreiber (ORCID: https://orcid.org/0000-0002-8827-1838)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/1331
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00