ISO/TS 10974 Compliance in MRI Safety: A Comparative Study of Electromagnetic Solvers for Predicting RF Heating in AIMDs
Motivation: Accurate RF heating prediction is crucial for patient safety and regulatory compliance. Goal(s): Provide an unbiased evaluation of ANSYS HFSS and Sim4Life in predicting RF-induced heating of AIMDs. Approach: Simulation setups of a commercial coil were optimized in both platforms. The transfer function (TF) of a DBS lead was validated experimentally and used to predict in-vivo RF heating in clinically relevant lead trajectories in body models. Results: Both solvers demonstrated high accuracy in RF heating predictions, with no significant difference in mean temperature rise. ANSYS HFSS achieved faster simulation times (4.72 vs. 15.40 hours), suggesting greater computational efficiency. Impact: This study offers valuable insights into selecting a suitable electromagnetic solver for RF-safety evaluations of AIMDs, supporting regulatory compliance and clinical decision-making by comparing computational accuracy and efficiency.
Authors
- Shahid Hameed (ORCID: https://orcid.org/0000-0001-9190-3472)
- Laleh Golestanirad (ORCID: https://orcid.org/0000-0003-3869-6114)
- Fuchang Jiang (ORCID: https://orcid.org/0000-0003-4161-6269)
- Sana Ullah (ORCID: https://orcid.org/0000-0002-2200-2897)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/0181
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00