Non-Cartesian Imaging on a Head-Only System: Impact and Correction of Higher Order Eddy Currents
Motivation: Image quality from non-Cartesian methods on a head only scanner were severely degraded with artifacts suggesting uncorrected eddy currents, even though 0th and 1st order corrections were used. Goal(s): The aim is to measure and correct for higher order (2nd order, including cross term) eddy currents. Approach: A sequence was developed to measure the eddy currents and an efficient image reconstruction method was develop to account for the non-linear eddy currents. Results: The presence of second order cross-term eddy currents (xz and yz) was confirmed. Image quality was improved with the new reconstruction algorithm. Impact: Non-Cartesian imaging can now be performed on the head only MAGNUS MRI scanner without degradation due to non-linear eddy currents.
Authors
- Steven Kecskemeti (ORCID: https://orcid.org/0000-0003-3589-430X)
- Kevin M. Johnson (ORCID: https://orcid.org/0000-0002-5807-7866)
- Andrew L. Alexander (ORCID: https://orcid.org/0000-0003-2695-768X)
- James F. Holmes
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/1341
- Primary Topic
- Non-Destructive Testing Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00