A fast stroke imaging protocol at 0.5T using high performance gradients

Motivation: Fast imaging is required for stroke protocols, and mid-field is generally assumed to require longer scan times. Goal(s): Acquire a set of diffusion weighted, susceptibility weighted, and fluid attenuated inversion recovery images as rapidly as possible using a 0.5T head-only MRI equipped with high performance gradients. Approach: We used a set of protocols designed to take advantage of the reduced geometrical distortions, short T1, and long T2 available at mid-field. Results: A full set of diagnostic quality images was acquired in less than five minutes of scan time. Impact: A 0.5T head-only MRI equipped with high performance gradients can potentially be used for rapid diagnosis of acute stroke. The small footprint, low field, and head-only design makes this magnet easier to site, improving MRI access for acute stroke patients.

Authors

Publication Details

Journal
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Published
2025-09-16
DOI
https://doi.org/10.58530/2025/5061
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A fast stroke imaging protocol at 0.5T using high performance gradients

Andrew Curtis, Sofia Chavez, Jeff A Stainsby, Curtis N. Wiens et al.
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Advanced MRI Techniques and Applications
article

A fast stroke imaging protocol at 0.5T using high performance gradients

Andrew Curtis, Sofia Chavez, Jeff A Stainsby, Curtis N. Wiens, Chad Harris, Wendy Oakden
article en

Abstract

Motivation: Fast imaging is required for stroke protocols, and mid-field is generally assumed to require longer scan times. Goal(s): Acquire a set of diffusion weighted, susceptibility weighted, and fluid attenuated inversion recovery images as rapidly as possible using a 0.5T head-only MRI equipped with high performance gradients. Approach: We used a set of protocols designed to take advantage of the reduced geometrical distortions, short T1, and long T2 available at mid-field. Results: A full set of diagnostic quality images was acquired in less than five minutes of scan time. Impact: A 0.5T head-only MRI equipped with high performance gradients can potentially be used for rapid diagnosis of acute stroke. The small footprint, low field, and head-only design makes this magnet easier to site, improving MRI access for acute stroke patients.

Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Openalex Percentile: Top 63%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.