Concentric-loops coil reducing imaging noise and enhancing imaging SNR compared to traditional single-loop coil
Motivation: To improve MR imaging performance using advanced RF coil. Goal(s): To provide a new RF coil with largely improved signal-to-noise (SNR) compared to traditional single-loop surface coils. Approach: We designed a two concentric (primary and secondary) loops RF coil, and evaluated the coil performance via electromagnetic simulation and phantom tests. Results: The concentric-loops coil produces 1.4-1.6 times higher B1 close to the coil plane, and 15% to 30% noise reduction, resulting in approximately 1.7-2.6 folds SNR gain compared to the traditional single-loop coil across different nuclei and field strengths (17O at 10.5T, 31P at 7T and 1H at 1.5T). Impact: This simple novel coil can be optimized and adapted for a broad range of MRS and MRI imaging applications with largely reduced imaging noise, improved coil sensitivity and approximately doubled imaging SNR as compared to the traditional single-loop coil.
Authors
- Wei Chen (ORCID: https://orcid.org/0000-0002-9608-1444)
- Xin Li (ORCID: https://orcid.org/0000-0001-6999-0610)
- Xiao‐Hong Zhu (ORCID: https://orcid.org/0009-0008-5926-1869)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/4248
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00