Motivation: Mitigate shading in MRI images, resulting from inhomogeneities in the B1+ field, which can obscure anatomical details and affect diagnostic accuracy. Goal(s): Evaluate feasibility of using dynamically calculated spatially selective RF pulses aiming to mitigate B1+ shading in 3D turbo-spin-echo imaging of the brachial plexus at 3T. Approach: Nonselective rectangular RF pulses are replaced by online-optimized spatially selective RF pulses based on B1 and B0 maps acquired immediately before the actual measurement. Results: Spatial-selective RF pulses can mitigate image shading caused by B1+ variations in 3D turbo-spin-echo imaging of the brachial plexus at 3T. Impact: The work shows that sophisticated pulse design can mitigate B1+ shading at 3T particularly in 3D turbo-spin-echo imaging of the brachial plexus at the cost of about 35% increased pulse energy and 40s increased scan time due to additional preparations.
Authors
- David Grodzki
- Jürgen Herrler (ORCID: https://orcid.org/0000-0002-4620-8216)
- Nicolas Groß‐Weege (ORCID: https://orcid.org/0000-0003-3507-6980)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/4415
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00