Differentiable Image-Based Needle Localization in MR-Guided Interventions
Motivation: MR-guided needle interventions require accurate and precise knowledge of the needle position and orientation, which is limited by image distortions. Goal(s): Provide differentiable image-based needle localization that inherently leverages image distortions that typically produce positioning errors. Approach: Needle susceptibility maps were modeled with a mixture of Gaussian functions that are related differentiably to a simulated distorted image. The simulated image was iteratively compared to a measured image to recover the true needle position and orientation. Simulations validated the method compared to a PCA approach. Results: The proposed approach achieved up to 50% lower position and angular errors than the PCA approach. Impact: Accurate needle localization is key to the safety and efficacy of interventional MR procedures. The proposed differentiable image-based method recovers needle position and pose accurately, regardless of needle-induced image distortions.
Authors
- William A. Grissom (ORCID: https://orcid.org/0000-0002-3289-1827)
- Sarah Garrow
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/4844
- Primary Topic
- Radiomics and Machine Learning in Medical Imaging
- Type
- article
- Field-Weighted Citation Impact
- 0.00