Prediction and Retrospective Correction of 2D PC-MRI Background Phase Errors Using the Gradient Impulse Response Function
Motivation: Background phase errors caused by eddy currents and mechanical oscillations in PC-MRI are a significant source of measurement error that require a reliable correction method. Goal(s): To demonstrate the effectiveness of gradient impulse response function (GIRF) predictions for the retrospective correction of background phase errors in 2D-PCMRI. Approach: The gradient waveforms used to acquire 2D-PCMRI were convolved with GIRFs to predict background phase responses and correct the datasets. The correction is tested in static phantoms and volunteers (n=10). Results: Background phase errors were reduced by 75.7% in static phantom experiments and by 65.2% in the volunteer data. Impact: We tested a gradient impulse response function (GIRF) based prediction method to correct background phase errors in 2D PC-MRI data. Background phase errors were reduced by 75.7% in static phantom experiments and 65.2% in the volunteer data.
Authors
- Michael Loecher (ORCID: https://orcid.org/0000-0002-6533-1113)
- Daniel B. Ennis (ORCID: https://orcid.org/0000-0001-7435-1311)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/0049
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00