Probing Glymphatic Exchange Between CSF and Tissue Using Long-TE Arterial Spin Labeling: Simplified Model and Shorter Scan Time
Motivation: There is currently not a clinically feasible imaging technique for the evaluation of glymphatic function. Goal(s): We proposed a new framework to measure a simple index, CSF signal fraction in ASL, which only requires 10 minutes of scan time using a single delay and two echoes. Approach: Compared to standard pCASL, the proposed sequence used enhanced background suppression, complex control and label subtraction, and a T2-preparation module with a long effective TE(eTE). Results: The parametric map had a variability of approximately 5%. An eTE of ≥480ms and a PLD of ≥2500ms are suitable for CSF fraction mapping. Impact: In this study, we developed a framework to measure CSF signal fraction in ASL MRI, which may provide a physiological index to evaluate tissue CSF dynamics and brain waste clearance system.
Authors
- Wen Shi (ORCID: https://orcid.org/0000-0002-5312-4889)
- Dengrong Jiang (ORCID: https://orcid.org/0000-0002-3843-1020)
- Jie Song (ORCID: https://orcid.org/0000-0001-7554-5564)
- Hanzhang Lu (ORCID: https://orcid.org/0000-0003-3871-1564)
- Dinil Sasi Sankaralayam (ORCID: https://orcid.org/0000-0002-6669-5584)
- Yulin Ge (ORCID: https://orcid.org/0000-0002-2320-6031)
- Yifan Gou (ORCID: https://orcid.org/0009-0009-5836-0299)
- Zihan Wang (ORCID: https://orcid.org/0009-0006-0531-8669)
- Zhiyi Hu (ORCID: https://orcid.org/0009-0002-6222-8126)
- Hongli Fan (ORCID: https://orcid.org/0000-0002-7938-1875)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/0251
- Primary Topic
- Cerebrospinal fluid and hydrocephalus
- Type
- article
- Field-Weighted Citation Impact
- 0.00