Feasibility of Water-Fat Separation for 2D Cardiac CINE Imaging with a Phase-Sensitive Approach on a 0.6T magnet

Motivation: Both fat and water appear bright in a bSSFP cine CMR. Distinguishing fat and water signals will provide additional information for diagnosis. However, fat-water separation becomes more difficult on a midfield scanner. Goal(s): To achieve fat-water separation without additional time cost on a 0.6T scanner. Approach: A phase sensitive algorithm was implemented to separate fat signals from a bSSFP EPI cine CMR scan. Results: The bSSFP EPI sequence enables shorter scan time and lower SAR than the conventional sequence. In addition, fat and water signals appear out-of-phase in the images and become separable with the phase sensitive algorithm after background phase correction. Impact: Water and fat signals become difficult to separate with midfield MRI scanners. The presented work utilizes an EPI scan to improve scan speed of a cine series while achieving fat and water separation with the phase information.

Authors

Publication Details

Journal
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Published
2025-09-16
DOI
https://doi.org/10.58530/2025/0323
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Feasibility of Water-Fat Separation for 2D Cardiac CINE Imaging with a Phase-Sensitive Approach on a 0.6T magnet

Jouke Smink, Tzu Cheng Chao, Dinghui Wang, Ece Ercan et al.
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Advanced MRI Techniques and Applications
article

Feasibility of Water-Fat Separation for 2D Cardiac CINE Imaging with a Phase-Sensitive Approach on a 0.6T magnet

Jouke Smink, Tzu Cheng Chao, Dinghui Wang, Ece Ercan, Tim Leiner, Spencer Waddle, Jacinta E. Browne, Michal Považan
article en

Abstract

Motivation: Both fat and water appear bright in a bSSFP cine CMR. Distinguishing fat and water signals will provide additional information for diagnosis. However, fat-water separation becomes more difficult on a midfield scanner. Goal(s): To achieve fat-water separation without additional time cost on a 0.6T scanner. Approach: A phase sensitive algorithm was implemented to separate fat signals from a bSSFP EPI cine CMR scan. Results: The bSSFP EPI sequence enables shorter scan time and lower SAR than the conventional sequence. In addition, fat and water signals appear out-of-phase in the images and become separable with the phase sensitive algorithm after background phase correction. Impact: Water and fat signals become difficult to separate with midfield MRI scanners. The presented work utilizes an EPI scan to improve scan speed of a cine series while achieving fat and water separation with the phase information.

Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Openalex Percentile: Top 63%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.