In vivo Glx measurements from GABA-edited HERMES at 3T are not consistent with those from short-TE PRESS across scanners, regions, and age groups.
Motivation: The reliability of 80ms HERMES for quantification of Glx (glutamate + glutamine) has not been assessed. Goal(s): Investigate the agreement between paired Glx measurements from GABA-edited HERMES (TE = 80ms) and conventional PRESS (TE = 35ms). Approach: HERMES and PRESS were acquired from two voxels (anterior cingulate cortex and thalamus) across 272 participants, 4 scanners, 2 age groups and 2 diagnostic groups. Results: We observe poor agreement between HERMES and PRESS Glx measurements, with consistent systematic and proportional differences between measurements across all scanners, age groups and diagnostic groups. Impact: In a large sample, we show that HERMES (TE = 80ms) Glx estimates are not consistent with paired Glx measurements from short-TE PRESS, highlighting the need for consideration during MRS sequence selection, integrating data across different acquisitions, and data interpretation.
Authors
- Herbert Roeyers (ORCID: https://orcid.org/0000-0002-4895-583X)
- Declan Murphy (ORCID: https://orcid.org/0000-0002-6664-7451)
- Pascal-M. Aggensteiner (ORCID: https://orcid.org/0000-0002-1048-9044)
- Tomoki Arichi (ORCID: https://orcid.org/0000-0002-3550-1644)
- Nicolaas A.J. Puts (ORCID: https://orcid.org/0000-0003-1024-1927)
- Terje Falck‐Ytter (ORCID: https://orcid.org/0000-0001-9714-0197)
- Eva Loth (ORCID: https://orcid.org/0000-0001-9458-9167)
- Alice Thomson
- Jan Buitelaar
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/0596
- Primary Topic
- Molecular Biology Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00