How close are we to the theoretical maximum SNR in portable low field MRI?
Motivation: Many groups build their own low field scanners, with different component designs. Relative performances have not been well documented due to a lack of detailed specifications and tests. Goal(s): To provide a set of measurements to benchmark the performance of low field scanners. Approach: The theoretical signal is calculated from first principles, including coil Q-value and transmit efficiency. Experimental data is acquired using different coils/samples and compared with the theoretical value. The effect of external noise sources and sub-optimal coil construction is also considered. Results: Experimental data agree well with theoretical predictions both for "optimal" and deliberately sub-optimal conditions. Impact: A framework has been established to estimate the signal-to-noise ratio obtainable on any given design of point-of-care low field imaging system.
Authors
- Chloé Najac (ORCID: https://orcid.org/0000-0002-7804-2281)
- Andrew Webb (ORCID: https://orcid.org/0000-0003-4045-9732)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/0490
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00