METHOD for CENTER FREQUENCY CALIBRATION USING STIR ON/OFF SPECTRA

Motivation: To propose a robust method for center frequency (F0) calibration. Goal(s): To improve image quality across a wide range of clinical conditions by improving the accuracy and robustness of center frequency calibration. Approach: A model-based peak fitting approach is applied to spectra acquired with STIR fat suppression on and off. The model kernels are tuned individually for STIR on/off data. By jointly solving F0 in both STIR on/off data the F0 calibration is robust even in the presence of strong fat signal (e.g. breast). Results: The proposed method was successful in 100% of test cases (51/51) across a range of human anatomies. Impact: Accurate and consistent center frequency calibration is critical for reliable image quality. The proposed method is robust to a wide range of clinical conditions and anatomies, including tissues with high off-resonance (e.g. cervical spine) or high fat content (e.g. breast).

Authors

Publication Details

Journal
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Published
2025-09-16
DOI
https://doi.org/10.58530/2025/3806
Primary Topic
Sensor Technology and Measurement Systems
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

METHOD for CENTER FREQUENCY CALIBRATION USING STIR ON/OFF SPECTRA

Andrew Wheaton
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Sensor Technology and Measurement Systems
article

METHOD for CENTER FREQUENCY CALIBRATION USING STIR ON/OFF SPECTRA

Andrew Wheaton
article en

Abstract

Motivation: To propose a robust method for center frequency (F0) calibration. Goal(s): To improve image quality across a wide range of clinical conditions by improving the accuracy and robustness of center frequency calibration. Approach: A model-based peak fitting approach is applied to spectra acquired with STIR fat suppression on and off. The model kernels are tuned individually for STIR on/off data. By jointly solving F0 in both STIR on/off data the F0 calibration is robust even in the presence of strong fat signal (e.g. breast). Results: The proposed method was successful in 100% of test cases (51/51) across a range of human anatomies. Impact: Accurate and consistent center frequency calibration is critical for reliable image quality. The proposed method is robust to a wide range of clinical conditions and anatomies, including tissues with high off-resonance (e.g. cervical spine) or high fat content (e.g. breast).

Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Openalex Percentile: Top 71%
Sensor Technology and Measurement Systems
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.