An inherently decoupled three-element solenoidal array coil for accelerated imaging on a 46 mT Halbach-based permanent magnet system.
Motivation: Accelerate wrist imaging on low field point of care MRI. Goal(s): Introducing a solenoidal phased array coil to reduce scan time with compressed sensing. Approach: Using electrical and geometrical decoupling methods to create three inherently decoupled solenoidal phased array and accelerating scan by using compressed sensing. Results: Inter-element coupling, and S-parameters for each of the three individual Rx coils and the transmit coil under loaded conditions were lower than -25 dB. A 3x compressed sensing acceleration produced results comparable to fully sampled k-space. Impact: This method of decoupling can make the application of solenoidal phase array possible for wrist imaging accelerate by factor 3 which is very important in point of care MRI systems.
Authors
- Yiming Dong (ORCID: https://orcid.org/0000-0002-8580-4555)
- Javad Parsa (ORCID: https://orcid.org/0000-0002-0845-8492)
- Andrew Webb (ORCID: https://orcid.org/0000-0003-4045-9732)
- Beatrice Lena
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/0498
- Primary Topic
- Atomic and Subatomic Physics Research
- Type
- article
- Field-Weighted Citation Impact
- 0.00