Diffusion MRI with a Multi-Shot Rosette Readout
Motivation: Demonstrate image quality of multi-shot rosette acquisitions in diffusion-weighted MRI (dMRI) in both ex vivo and in vivo applications. Goal(s): Use redundant k-space sampling of rosette readouts to measure and correct for magnetic fields (B0 and shot-to-shot phase variations) that typically cause distortions in dMRI acquisitions Approach: Rosette dMRI acquisitions were evaluated on ex vivo ferret brains and in vivo mouse spinal cords. Results: Incorporating B0 and shot-to-shot phase variations estimated from rosette acquisitions without the need prescans showcases the potential of rosette readouts in advancing dMRI image quality. Impact: This abstract demonstrates multi-shot rosette readouts in diffusion MRI, including using redundant k-space information for estimating B0 and phase variations to improve image quality, evaluated in ex vivo ferret brains and in vivo mouse spinal cord.
Authors
- Mark D. Does (ORCID: https://orcid.org/0000-0003-2701-6208)
- Kevin D. Harkins (ORCID: https://orcid.org/0000-0003-3579-9273)
- Jonathan B. Martin (ORCID: https://orcid.org/0000-0002-9384-8056)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/0517
- Primary Topic
- Advanced Neuroimaging Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00