Effects of Implantable Pulse Generator (IPG) Impedance on the MRI Conditionality of Active Implantable Medical Devices (AIMDs) under 1.5T MRI
Motivation: To understand the interchangeability the IPGs for an AIMD system under MR RF exposure. Goal(s): To evaluate/compare the effect of AIMD input impedance on both the RF-induced heating and the RF-induced voltage under RF exposure. Approach: Evaluate the AIMD models for both RF-induced heating/voltage using an implantable lead connected to IPGs with different input impedances. The AIMD models (transfer functions) are evaluated via the reciprocity approach. Results: When the lead characteristic impedance and IPG impedance differ greatly, the device heating stays consistent. However, the induced voltage changes. Both RF-induced heating/voltage are crucial for the interchangeability of IPGs for an AIMD system. Impact: This study emphasizes the significance of on the RF-induced voltage at the IPG in addition to heating at the distal end. The analysis method presented could serve as a valuable approach for assessing the interchangeability of IPGs for AIMDs.
Authors
- Hongbae Jeong (ORCID: https://orcid.org/0000-0003-4908-2070)
- Farshad Ebrahimi
- Jianfeng Zheng (ORCID: https://orcid.org/0000-0001-8998-9561)
- Qingyan Wang (ORCID: https://orcid.org/0000-0001-6104-4048)
- Ji Chen (ORCID: https://orcid.org/0000-0001-6944-2729)
- Ananda Kumar
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/3956
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00