DIM: A Diffusion Instability Measure
Motivation: QA in diffusion MRI is challenging and rarely implemented. We provide a new QA measure that is borrowed from fMRI. Goal(s): To test a diffusion instability measure (DIM) on HARDI data from different scanners, with and without concurrent magnetic field monitoring (FM). Approach: We estimate the correlation coefficient matrix between all pairs of diffusion directions across voxels inside the phantom and compute a QA measure from the eigenvalues of the matrix. Results: DIM is sensitive to data quality affected by b-values, TE, acquisition/reconstruction types. Data acquired with FM showed a DIM value that was, on average, 10 times smaller than those without FM. Impact: We developed and tested a diffusion instability measure (DIM) that is sensitive to factors affecting image quality, such as magnetic field monitoring, f0 stabilization, TE, gradient strengths, making it a promising tool for routine diffusion MRI quality assurance.
Authors
- Zoltán Nagy (ORCID: https://orcid.org/0000-0001-6611-6362)
- Tim Schmidt (ORCID: https://orcid.org/0000-0003-0135-6335)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/4049
- Primary Topic
- Advancements in Photolithography Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00