Accurate assessment of 11.7 T head coil energy deposition using field-drift tolerant proton resonance frequency-based MR thermometry
Motivation: RF transmit safety for high-field MRI. Goal(s): Validation of RF heating simulations for a human head transmit/receive coil for 11.7 T using high-sensitivity field-drift tolerant MR thermometry. Approach: Isopropanol was added to a polyvinylpyrrolidone head phantom to serve as an internal reference for MR thermometry. This phantom was used to compare the RF energy deposition by a 500 MHz head coil with FDTD simulations. Results: Temperature precision better than 20 mK was achieved. Correspondence between temperature maps and SAR prediction from FDTD simulations demonstrated validity of the approach for high-field coil safety assessment. Impact: A state-of-the-art method for high-field transmit coil performance assessment is presented: High-accuracy FDTD simulation of SAR deposition from a 500 MHz human head coil was validated using a robust MR thermometry method, based on an internally-referenced phantom material.
Authors
- Natalia Gudino (ORCID: https://orcid.org/0000-0002-7729-4911)
- Peter van Gelderen (ORCID: https://orcid.org/0000-0001-8117-8270)
- Joe Murphy‐Boesch
- Jacco de Zwart
- Jeff H. Duyn
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/3959
- Primary Topic
- Nuclear Physics and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00