Active Electromagnetic Interference Suppression for Real-Time MR Thermometry During MR-Guided Microwave Ablation

Motivation: Electromagnetic interference (EMI) from the microwave system may severely disrupt MR thermometry during microwave ablation (MWA). While hardware modification to the MWA system has been proposed, it may compromise MWA effectiveness or complicate workflow. Goal(s): Develop a software-based EMI suppression technique for real-time MR thermometry during MWA. Approach: We designed a domain transfer function-based technique to directly estimate and remove EMI signals from acquired data. We tested its effectiveness in phantom experiments. Results: The proposed technique demonstrated up to 90% EMI suppression rate. Compared to temperature probes, real-time MR thermometry achieved mean-absolute-error of 1.30°C in the ablation zone during MWA. Impact: The software-based active EMI suppression technique enables reliable real-time MR temperature monitoring during continuous microwave ablation without additional hardware modifications. This technique has potential for clinical deployment to enhance patient safety and improve treatment outcomes.

Authors

Publication Details

Journal
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Published
2025-09-16
DOI
https://doi.org/10.58530/2025/0677
Primary Topic
Ultrasound and Hyperthermia Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Active Electromagnetic Interference Suppression for Real-Time MR Thermometry During MR-Guided Microwave Ablation

David S.K. Lu, Qing Dai, Jason Chiang, Shu‐Fu Shih et al.
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Ultrasound and Hyperthermia Applications
article

Active Electromagnetic Interference Suppression for Real-Time MR Thermometry During MR-Guided Microwave Ablation

David S.K. Lu, Qing Dai, Jason Chiang, Shu‐Fu Shih, Holden H. Wu
article en

Abstract

Motivation: Electromagnetic interference (EMI) from the microwave system may severely disrupt MR thermometry during microwave ablation (MWA). While hardware modification to the MWA system has been proposed, it may compromise MWA effectiveness or complicate workflow. Goal(s): Develop a software-based EMI suppression technique for real-time MR thermometry during MWA. Approach: We designed a domain transfer function-based technique to directly estimate and remove EMI signals from acquired data. We tested its effectiveness in phantom experiments. Results: The proposed technique demonstrated up to 90% EMI suppression rate. Compared to temperature probes, real-time MR thermometry achieved mean-absolute-error of 1.30°C in the ablation zone during MWA. Impact: The software-based active EMI suppression technique enables reliable real-time MR temperature monitoring during continuous microwave ablation without additional hardware modifications. This technique has potential for clinical deployment to enhance patient safety and improve treatment outcomes.

Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Openalex Percentile: Top 76%
Ultrasound and Hyperthermia Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.