Highly Accelerated Pulmonary MRI Using RF Depolarization of 129Xe: The Sectoral Pulse Sequence
Motivation: Lung MRI using hyperpolarized 129Xe is an important but expensive diagnostic tool for various pulmonary diseases. Reducing scan times and the xenon dose required for a clinically-viable lung image would drastically lower the barrier to entry for pulmonary imaging. Goal(s): We intend to use the RF depolarization of xenon to accelerate lung images and reduce the required amount of inhaled 129Xe. Approach: Various sampling schemes for the Sectoral pulse sequence are developed and compared with the gradient echo sequence. Results: The proposed method, coupled with the short-Sectoral sampling scheme, allows for aggressive undersampling with a third of a typical xenon dose. Impact: We present a significant acceleration method which uses RF depolarization of hyperpolarized gas in lungs to reduce 129Xe dose costs. This method results in clinically-viable lung images with up to 10-fold acceleration possible with a 3-fold reduction in 129Xe dose.
Authors
- Alexei Ouriadov (ORCID: https://orcid.org/0000-0002-0920-2528)
- Matthew C. Fox
- Samuel Perron
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/2593
- Primary Topic
- Atomic and Subatomic Physics Research
- Type
- article
- Field-Weighted Citation Impact
- 0.00