Flexible, twisted-pair coil array with dielectric resonators and transceiver dipole antennas for enhanced sensitivity in MRI at 7T
Motivation: Increasing SNR in MRI by designing flexible receive arrays that adapt closely to patient anatomy is needed to improve clinical image quality. Goal(s): This study aims to develop a flexible, multi-channel twisted-pair (TP) coil array with integrated dielectric resonators (DRAs) and transceiver dipole antennas to enhance SNR at 7T. Approach: The TP array, combined with DRAs and dipole antennas, was validated via numerical simulations and phantom experiments on a 7T MRI scanner to evaluate B1+ efficiency and SNR. Results: Integrating DRAs, dipole antennas and TP coils provided more than a 2-fold gain in peripheral SNR and almost a 1.2-fold gain in the center. Impact: This new approach can substantially boost receive performance of flexible RF coils tailored for various clinical applications.
Authors
- Irena Zivkovic (ORCID: https://orcid.org/0000-0001-6900-6580)
- Mark Widmaier (ORCID: https://orcid.org/0009-0000-3353-0305)
- Daniel Wenz (ORCID: https://orcid.org/0000-0002-3216-7599)
- Lijing Xin (ORCID: https://orcid.org/0000-0002-5450-6109)
- Jules Vliem (ORCID: https://orcid.org/0009-0008-6680-6345)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/3318
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00