On the source of unresolved Nyquist ghosting in EPI at high and ultrahigh field
Motivation: Residual Nyquist ghosting commonly confounds echo-planar images at ultrahigh field. The effect is not yet fully understood. Goal(s): To elucidate why standard ghost correction methods may fail at 7T. Approach: We analyzed the shift between odd and even echoes based on a linear, time-invariant (LTI) gradient model for different gradient axes and field strengths. In EPI experiments, we quantified the artifacts for corrections based on three navigators and the LTI-model. Results: Mechanical resonances of the gradient coils cause oscillating shifts between odd and even echoes. Through appropriate parameter choices or trajectory correction, they can be avoided or corrected for. Impact: Residual Nyquist ghosting in EPI at 7T can be reduced by avoiding or correcting for mechanical oscillations of the gradient coils. A deepened understanding of this effect allows for more reliable correction strategies and less confounded functional or diffusion-weighted MRI.
Authors
- Istvan Homolya
- Herbert Köstler (ORCID: https://orcid.org/0000-0001-6207-9226)
- Hannah Scholten (ORCID: https://orcid.org/0000-0002-9049-7668)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/3281
- Primary Topic
- Laser-Matter Interactions and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00