Respiratory motion-corrected model-based 3D water-fat MRA at 0.55T
Motivation: Navigator-gated MRAs often have unpredictable scan times, limiting practicality. Achieving robust fat suppression remains challenging at 0.55T. Goal(s): Develop a 5-minute MRA at 0.55T with predictable scan time, 100% data efficiency, and reliable water/fat separation, comparable to navigator-gated results. Approach: Use self-gating to infer motion bins and apply model-based Dixon reconstruction with motion correction for direct water/fat imaging. Compare results to navigator-gated acquisitions with image-based water-fat separation. Results: The proposed method offers comparable image quality, 100% efficiency, and predictable scan time, while avoiding navigator-induced saturation effects. Impact: An MRA at 0.55T covering the entire thorax, allowing water/fat separation without navigator artefacts can be achieved in a predictable time of 5 minutes.
Authors
- Daniel Giese (ORCID: https://orcid.org/0000-0002-3925-4441)
- Dominik M. Nickel (ORCID: https://orcid.org/0000-0002-0360-7233)
- Christoph Kolbitsch
- Michaela Schmidt (ORCID: https://orcid.org/0000-0003-0184-5171)
- Robert D. Stoll
- Tobias Schäffter
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/3671
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00