Evaluation of background phase errors from eddy currents and mechanical oscillations in phase contrast magnetic resonance imaging
Motivation: Background phase errors in PC-MRI include eddy currents and mechanical oscillations. Relative contributions over a wide range of velocity encoding gradients (VENCs) remain underexplored. Goal(s): To evaluate background phase errors across clinically relevant VENCs, characterizing the impact of eddy currents and mechanical oscillations. Approach: Waveforms with a range of VENCs were generated, and the gradient impulse response function was used to predict phase errors as a sum of eddy current and mechanical oscillation effects. Phantom experiments across different VENCs were conducted. Results: Eddy current effects dominated lower VENCs. Between approximately 52 and 92 cm/s the phase error was largely oscillatory. Impact: The necessary correction for background phase errors depends on the VENCs, which explains why errors are protocol dependent.
Authors
- Simon Thalén (ORCID: https://orcid.org/0000-0002-7046-5378)
- Michael Loecher (ORCID: https://orcid.org/0000-0002-6533-1113)
- Daniel B. Ennis (ORCID: https://orcid.org/0000-0001-7435-1311)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/3660
- Primary Topic
- Nuclear Physics and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00