A fast eddy currents correction method based on field probe in portable low-field MRI

Motivation: Significant field inhomogeneity, more pronounced residual magnetism of supporting structure, and field drift over time of portable systems preclude accurate eddy currents measurement. Goal(s): To propose a fast eddy currents measurement and correction method based on field probe, tailored to the characteristics of portable low-field systems. Approach: A frequency correction term was introduced in the probe localization sequence. The testing gradients and fitting algorithms for the eddy currents measurement were optimized. The correction quality was evaluated in vivo with diffusion-weighted EPI. Results: The calibration took approximately 5min, reducing eddy currents to below 0.03% and effectively minimizing distortions in DW-EPI. Impact: Eddy currents in a samarium cobalt two-plate permanent magnet portable design can be reliably characterized in 5 minutes, which can be used to minimize impacts of field inhomogeneity, instability, and remanence, as well as improving clinical image quality (e.g., DWI).

Authors

Publication Details

Journal
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Published
2025-09-16
DOI
https://doi.org/10.58530/2025/0493
Primary Topic
Non-Destructive Testing Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A fast eddy currents correction method based on field probe in portable low-field MRI

Suen Chen, Shiwei Yang, Philip R. Lee, Yueqi Qiu et al.
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Non-Destructive Testing Techniques
article

A fast eddy currents correction method based on field probe in portable low-field MRI

Suen Chen, Shiwei Yang, Philip R. Lee, Yueqi Qiu, Hao Chen, Zhiyong Zhang
article en

Abstract

Motivation: Significant field inhomogeneity, more pronounced residual magnetism of supporting structure, and field drift over time of portable systems preclude accurate eddy currents measurement. Goal(s): To propose a fast eddy currents measurement and correction method based on field probe, tailored to the characteristics of portable low-field systems. Approach: A frequency correction term was introduced in the probe localization sequence. The testing gradients and fitting algorithms for the eddy currents measurement were optimized. The correction quality was evaluated in vivo with diffusion-weighted EPI. Results: The calibration took approximately 5min, reducing eddy currents to below 0.03% and effectively minimizing distortions in DW-EPI. Impact: Eddy currents in a samarium cobalt two-plate permanent magnet portable design can be reliably characterized in 5 minutes, which can be used to minimize impacts of field inhomogeneity, instability, and remanence, as well as improving clinical image quality (e.g., DWI).

Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Openalex Percentile: Top 75%
Non-Destructive Testing Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.