Evaluation of Pediatric MRI with a High-Density Posterior Cushion Coil
Motivation: Posterior coil arrays built into MRI tables are common, but provide suboptimal performance in pediatric imaging due to their distance from the imaged volume and lower coil density. Goal(s): Improve pediatric MRI by constructing a prototype 30 channel high density posterior cushion coil (HDPCC) incorporated into a padded cushion. Approach: 25 pediatric patients were scanned with HDPCC. 25 age-matched controls were scanned with standard coils. Two radiologists graded image quality on a 5-point scale. SNR was measured in the right paraspinal musculature and the liver. Results: All images obtained with HDPCC were diagnostic (mean score 4.13/5). Posterior SNR was higher than standard coils. Impact: HDPCC can improve SNR in pediatric MRI by conforming to the body and minimizing distance to the imaged volume.
Authors
- John M. Pauly (ORCID: https://orcid.org/0000-0001-5918-4172)
- Yun Jeong Stickle
- Jana Vincent (ORCID: https://orcid.org/0000-0002-2707-6000)
- Praveen Jayapal (ORCID: https://orcid.org/0000-0003-2869-1432)
- Mark Giancola
- Clyve Follante
- Thomas Grafendorfer
- Fraser Robb (ORCID: https://orcid.org/0000-0002-1625-9709)
- Shreyas Vasanawala (ORCID: https://orcid.org/0000-0002-1999-6595)
- Terrence Jao (ORCID: https://orcid.org/0000-0001-6951-1894)
- Greig Scott (ORCID: https://orcid.org/0000-0002-6609-5431)
- Ali Syed
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/0538
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00