The MR safety scan condition requiring for the external fixator to stay outside the scanner bore should be revised

Motivation: MRI scans for patients with external fixator, even made of non-ferromagnetic material, are often prohibited because the scan conditions given by the manufacturer would state the external fixator should stay outside the scanner bore. Goal(s): This study was performed to revisit this MRI safety scan condition and help some of the patients with ex-fix device in their clinical management. Approach: This study was performed with a phantom test at a 3T scanner. Results: The results suggest that the possible RF-induced heating is negligible if the ex-fix device stays outside a certain range from the isocenter but still inside the MRI scanner bore. Impact: The results of this study strongly suggests the patients with external fixator, who have often been prevented from getting MRI scans, could be allowed to get MRI scans, if the external fixator is at certain distance away from the isocenter.

Authors

Publication Details

Journal
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Published
2025-09-16
DOI
https://doi.org/10.58530/2025/3967
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The MR safety scan condition requiring for the external fixator to stay outside the scanner bore should be revised

Atsushi Takahashi, Jacob S.W. Borgida, Seonghwan Yee, Thuan V Ly et al.
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Advanced MRI Techniques and Applications
article

The MR safety scan condition requiring for the external fixator to stay outside the scanner bore should be revised

Atsushi Takahashi, Jacob S.W. Borgida, Seonghwan Yee, Thuan V Ly, Jad Husseini
article en

Abstract

Motivation: MRI scans for patients with external fixator, even made of non-ferromagnetic material, are often prohibited because the scan conditions given by the manufacturer would state the external fixator should stay outside the scanner bore. Goal(s): This study was performed to revisit this MRI safety scan condition and help some of the patients with ex-fix device in their clinical management. Approach: This study was performed with a phantom test at a 3T scanner. Results: The results suggest that the possible RF-induced heating is negligible if the ex-fix device stays outside a certain range from the isocenter but still inside the MRI scanner bore. Impact: The results of this study strongly suggests the patients with external fixator, who have often been prevented from getting MRI scans, could be allowed to get MRI scans, if the external fixator is at certain distance away from the isocenter.

Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Openalex Percentile: Top 64%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.