Impact of RF Coil Topology on Population Studies of RF Heating of Leaded Active Implants During MRI
Motivation: High-pass RF coil models are frequently used in numerical exposure assessments for implant safety, whereas many commercial 3T systems use band-pass coil topologies. Goal(s): To study the sensitivity of RF heating predictions on RF coil topology. Approach: High- and band-pass coil model simulations are performed and the tangential E-fields and deposited power predictions for generic implants compared for ViP model Fats at 3T. Results: Variation from lateral/dorsal displacement of Fats, different coil design or current feeding, or the use of a Huygens' box with a loaded or unloaded coil lie fully within the range of coverage of the high pass coil library. Impact: The sensitivity evaluation of Tier 3 studies to birdcage topology and simulation parameters indicates that at 3T, the birdcage topology does not contribute significant variation to Tier 3 applications compared to factors such as patient anatomy, imaging landmarks, and shimming.
Authors
- Tolga Goren (ORCID: https://orcid.org/0000-0002-4377-0676)
- Niels Kuster (ORCID: https://orcid.org/0000-0002-5827-3728)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/3970
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00