Optimizing the white matter/grey matter contrast on a portable 46 mT MRI scanner
Motivation: Low-field MRI, known for its accessibility and portability, faces challenges in distinguishing grey matter (GM) from white matter (WM) due to lower signal-to-noise ratios and converging relaxation times. Goal(s): This study aimed to improve WM/GM contrast in low-field MRI using a T1-weighted sequence with short inversion time (TI). Approach: T1-weighted sequence with optimized TI (90ms) was used to enhance contrast maintaining good SNR. Other TIs and resolution were also tested. Results: Results showed consistently high WM/GM contrast across subjects and resolution, with an optimal TI of 90 ms balancing contrast and SNR. Impact: This study demonstrates that a T1-weighted sequence with optimized inversion time effectively enhances grey-to-white matter contrast in low-field MRI, offering reliable tissue differentiation for different subjects and resolutions.
Authors
- Beatrice Lena (ORCID: https://orcid.org/0000-0001-9283-5918)
- Chloé Najac (ORCID: https://orcid.org/0000-0002-7804-2281)
- Thomas O’Reilly (ORCID: https://orcid.org/0000-0002-4547-2955)
- Andrew Webb (ORCID: https://orcid.org/0000-0003-1304-0434)
- Ruben van den Broek
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/4238
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00